Dialysis.c

Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
Handle CAN and UART error/warning notifications. Differentiated HD from DG s/w fault. Updated some older module/function headers to latest format.

  1. … 24 more files in changeset.
Bamboo Commit: Updated the Copyright with the current year and replaced tabs with 4 spaces

  1. … 70 more files in changeset.
Changed UF reference volume update approach from time x rate equation to accumulation of step time x rate. Fixed debug logging of event data when alarm is triggered. Updated flow test build output.

  1. … 4 more files in changeset.
Moved modes and sub-modes enums to common header. Added sub-mode to HD op mode broadcast. Added current UF rate to UF rate change response to UI.

  1. … 29 more files in changeset.
Increased target fill volume. Addressed code review feedback.

  1. … 2 more files in changeset.
Added test code that can read slow page of FPGA. Added test code for flow signal strength and occlusion sensor debug output for h/w testing. Delayed reservoir switch until after DG has flushed lines and started recirculating in order to get a more stable reading on load cells.

  1. … 7 more files in changeset.
Addressed code review feedback.

Misc UF fixes. Changed CAN from FIQ to IRQ to see if helps with intermittent lockup.

  1. … 10 more files in changeset.
Updated dialysis sub-mode to handle two reservoirs for UF.

  1. … 5 more files in changeset.
Merge branch 'master' into DEN-2390-hd-dialysate-temperature-1-of-2

  1. … 2 more files in changeset.
Updates to UT/IT. Minor code fixes per UT/IT.

  1. … 4 more files in changeset.
Code review fixes.

  1. … 2 more files in changeset.
Code review fixes. Moved task interrupts to notifications.

  1. … 6 more files in changeset.
Updated the Copyright with the current year and replaced tabs with 4 spaces

  1. … 6 more files in changeset.
Fixes per unit testing. Added UT/IT for NVDataMgmt module. Added UT for ModeTreatment module. Updated other unit tests.

  1. … 4 more files in changeset.
DEN-1778 Added alarms to NVDatMgmt and created a unit test in VectorCAST

  1. … 2 more files in changeset.
Split UF completed and off states.

  1. … 1 more file in changeset.
Updated alarm thresholds.

  1. … 2 more files in changeset.
Updated UF and Treatment time changes during treatment.

  1. … 2 more files in changeset.
Added messaging for blood/dialysate flow rate change and HD to DG dialysate temperature set points.

  1. … 9 more files in changeset.
Fixes per UT and IT. Updates to UT and IT.

  1. … 7 more files in changeset.
Updated UT for SystemCommMessages.

Updated doxygen comments. Un-macro test functions and override variables in DialOutFlow module.

  1. … 4 more files in changeset.
Updates to time and UF change messaging.

  1. … 2 more files in changeset.
Fixes for UF and treatment time changes. Updated pump factors.

  1. … 9 more files in changeset.
Further implementation of UF settings change messaging and handling. Code review fixes.

  1. … 17 more files in changeset.
Changed breadboard build switch to be more biased toward the new HD board. Added UF pause resume functions and message handling. Updated internal ADC voltage and conversion factors for new HD board. Commented out RTC POST until RTC on new HD board is configured.

  1. … 12 more files in changeset.
Put all debug UART code inside a debug build switch. Misc. other minor changes.

  1. … 19 more files in changeset.
Added IT for Dialysis module. Minor updates.

  1. … 4 more files in changeset.
Updated more doxygen comments.

  1. … 9 more files in changeset.