Matched DG arrangement and fix for CAN Tx. Added inlet conductivity alarm ID.

  1. … 5 more files in changeset.
Handle CAN and UART error/warning notifications. Differentiated HD from DG s/w fault. Updated some older module/function headers to latest format.

  1. … 21 more files in changeset.
Bamboo Commit: Updated the Copyright with the current year and replaced tabs with 4 spaces

  1. … 56 more files in changeset.
UT updates. Added UT for DGInterface. Misc fixes per UT.

  1. … 4 more files in changeset.
Changed UF reference volume update approach from time x rate equation to accumulation of step time x rate. Fixed debug logging of event data when alarm is triggered. Updated flow test build output.

  1. … 4 more files in changeset.
Added misc test/debug build flags. Added persistence to CAN buffer overflow faults so that we can try to recover from one-off spurts when a node is coming online.

  1. … 8 more files in changeset.
Fixed UF change verification logic. Fixed doxygen comments.

  1. … 3 more files in changeset.
Made reservoir target volumes public. Set all UF volumes to F32 type for consistency.

  1. … 4 more files in changeset.
Addressed code review feedback. Fixed treatment time minimum range to UI message.

  1. … 1 more file in changeset.
Fixed message payload sizing issue on UF option confirmation message handling.

  1. … 1 more file in changeset.
Changed max dialysate volume from 180 to 150 L. Split UF change response messages into 2 separate messages to prevent UI confusion.

  1. … 2 more files in changeset.
Added safety shutdown override command support. Zero treatment duration param in response to UF adjustment message if time adjustment should be disabled.

  1. … 6 more files in changeset.
Moved modes and sub-modes enums to common header. Added sub-mode to HD op mode broadcast. Added current UF rate to UF rate change response to UI.

  1. … 15 more files in changeset.
Increased target fill volume. Addressed code review feedback.

  1. … 1 more file in changeset.
Added test code that can read slow page of FPGA. Added test code for flow signal strength and occlusion sensor debug output for h/w testing. Delayed reservoir switch until after DG has flushed lines and started recirculating in order to get a more stable reading on load cells.

  1. … 7 more files in changeset.
Addressed code review feedback.

Added build switch to run UF test with or without DG.

  1. … 2 more files in changeset.
Misc UF fixes. Changed CAN from FIQ to IRQ to see if helps with intermittent lockup.

  1. … 8 more files in changeset.
Updated dialysis sub-mode to handle two reservoirs for UF.

  1. … 4 more files in changeset.
Implemented some test code of UF demo support in Standby mode. Started a state machine for treatment mode reservoir management.

  1. … 3 more files in changeset.
Added minimum 1 hr treatment to settings adjustment range checks. Changed blood/dialysate inlet pump flow rate filter approach. Changed blood/dialysate inlet pump control interval to 10 sec.

  1. … 2 more files in changeset.
Added pump alarms and checks.

  1. … 9 more files in changeset.
Merged RTC and NVData modules and associated UT/IT. UT/IT updates.

  1. … 2 more files in changeset.
UT/IT updates.

  1. … 2 more files in changeset.
DEN-1778 Merged master into the branch

  1. … 11 more files in changeset.
Added DG messaging.

  1. … 4 more files in changeset.
Added HD op mode broadcast.

  1. … 2 more files in changeset.
Merge branch 'master' into DEN-2390-hd-dialysate-temperature-1-of-2

  1. … 1 more file in changeset.
Fixes for eval board build for Behrouz.

  1. … 5 more files in changeset.
Updates to UT/IT. Minor code fixes per UT/IT.

  1. … 3 more files in changeset.