DENBUG-302 added UF rate events for information

DENBUG-298 updated the compatibility number

DENBUG-302 added a new generic command

DENBUG-298 changed the alarm ranks to 110 from 10 that are not fault anymore. Created a new reject reason for saline bolus not being allowed if the alarm does not allow blood recirc or dialysate recirc

DENBUG-308 made alarms 38 and 42 recoverable

DENBUG-298 changed some of the HD alarm from fault to be recoverable

DENBUG-308 updated the alarm properties of alarm 42, dialout rotor speed to be like alarm 38, dialin rotor speed

DENBUG-299 removed alarm 273

DENBUG-299 removed alarm 273 for release 21

DENBUG-298 removed the maximum additional saline bolus reject reason

DENBUG-298 removed the saline bolus max volume state and its reject reason

DENBUG-298 reverted the compatibility for testing

DENBUG-298 added a new state in treatment stop to support saline bolus

DENBUG-311 changed the properties of alarm 38

DENBUG-304 removed the newly added messages

DENBUG-304 added HD to DG RTC sync request and response messages

DENBUG-299: added alarm 273 for occlusion in prime

DENBUG-288 changed the alarm properties of alarms 147 and 129

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Merged DEN-17208

DEN-17208 updated the software compatibility

Merged DEN-17135

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Merged DEN-17135

Merge branch 'DEN-17135-concentrate-pump-calibration-staging' of bitbucket.org:diality-cloud/common into DEN-17135-concentrate-pump-calibration-staging

DENBUG-290 updated the alarms table

DENBUG-288 changed the new ranks of alarms 147 and 129 to the previous status

DENBUG-288 updated the alarm properties of alarm 129 and 147

DENBUG-290 updated the alarm properties of the alarm 129

DEN-17135 added an alarm for the concentrate pumps calibration record