Please add D99 temp average.

Please add D99 temp average.

Implement the filter logics using these variables.

Implement the filter logics using these variables.

50

50

Does this function need to support the rest of the level sensors? (e.g. P25, D63, D98?)

Does this function need to support the rest of the level sensors? (e.g. P25, D63, D98?)

Please add up bicarb chamber level based on the D63 and D98 level sensor combinations?

Please add up bicarb chamber level based on the D63 and D98 level sensor combinations?

D98?

D98?

This should be D98 function.

This should be D98 function.

These are redundant information from level and pressure module, we may remove them.

These are redundant information from level and pressure module, we may remove them.

Bicarb Chamber Low level sensor

Bicarb Chamber Low level sensor

fresh dialysate temperature?

fresh dialysate temperature?

freshDialysatePressurePostUltrafilter (instead of TMP)?

freshDialysatePressurePostUltrafilter (instead of TMP)?

Please remove it.

Please remove it.

RTDReadCnt -> Should be defined here?

RTDReadCnt -> Should be defined here?

Please remove RTDReadCnt, per HDD.

Please remove RTDReadCnt, per HDD.

remove extra blank line

remove extra blank line

LDT-2185 iofp alpha updates + review comment implementation

0.0 is OK

0.0 is OK

BiCarb related state machines will be called in pre-gen, gen dialysate and post-gen modes. currently its kept here, so any modes can set a flag to initiate the required operations.

BiCarb related state machines will be called in pre-gen, gen dialysate and post-gen modes. currently its kept here, so any modes can set a flag to initiate the required operations.

Done.

Done.

Done.

Done.

Updated.

Updated.

Done.

Done.

Done.

Done.

Done.

Done.

We will move towards 200 RPM being the minimum speed with updated Maxon controller. So, there is no need to maintain two different minimum RPM going forward.

We will move towards 200 RPM being the minimum speed with updated Maxon controller. So, there is no need to maintain two different minimum RPM going forward.

Done.

Done.

Done.

Done.

Use macro for 0.0F

Use macro for 0.0F