This doesn't look like a valid substate. We should have an enum for valid substates. In default case, we should trigger a s/w fault and also set the state to a valid state (idle or something like t...

This doesn't look like a valid substate. We should have an enum for valid substates. In default case, we should trigger a s/w fault and also set the state to a valid state (idle or something like that).

I think /// was correct. // will not be seen by doxygen.

I think /// was correct. // will not be seen by doxygen.

To make it exclusive operation. no previous request should be pending when a new operation is requested. So only one state machine is running at a time. otherwise, while chamber fill is in progress...

To make it exclusive operation. no previous request should be pending when a new operation is requested. So only one state machine is running at a time. otherwise, while chamber fill is in progress, anyone can request drain and both requests will be handled. even through new code run only one state machine at a time, this will ensure that only one request is submitted to drybicart

Not clear, why we are checking fill and supply request here?

Not clear, why we are checking fill and supply request here?

we should use the functions defined in testsupport.c/h -> getU32OverrideValue, getF32OverrideValue based on data type. this needs to be fixed throughout the file.

we should use the functions defined in testsupport.c/h -> getU32OverrideValue, getF32OverrideValue based on data type. this needs to be fixed throughout the file.

space after if. please fix it throughout the files.

space after if. please fix it throughout the files.

I guess, Fill and Drain are priority items than the Supply request, which means upon request of fill and drain, state machine should service that operation. Also, setting supply request should be ...

I guess, Fill and Drain are priority items than the Supply request, which means upon request of fill and drain, state machine should service that operation.

Also, setting supply request should be based on the level sensor later (there shouldn't be any default request -> currently supply is assigned)

Please introduce test config for drybicart selection.

Please introduce test config for drybicart selection.

Done.

Done.

Done

Done

Since trimmer heater is enabled, no need of temp compensation. only for testing purposes, temp compensation is needed. Hence by default, temp compensation is disabled.

Since trimmer heater is enabled, no need of temp compensation. only for testing purposes, temp compensation is needed. Hence by default, temp compensation is disabled.

For spent chamber fill, we don't know how low the liquid level is ( in Chamber H) and hence decided not to turn on the trimmer heater for safety reasons. For bicarb fill, the dry bicart code should...

For spent chamber fill, we don't know how low the liquid level is ( in Chamber H) and hence decided not to turn on the trimmer heater for safety reasons. For bicarb fill, the dry bicart code should cover the actuators status. the current bicarb chamber fill code here should be updated with the newly implemented dry bi cart code changes.

Fixed it

Fixed it

Remove extra blank line.

Remove extra blank line.

It is your comment, so please resolve it.

It is your comment, so please resolve it.

Merged LDT-2030 and resolved the conflicts

  1. … 6 more files in changeset.
It is replied by Vinay , there is no need to maintain two different rpm. so we can close this.

It is replied by Vinay , there is no need to maintain two different rpm. so we can close this.

It will be fixed in Balancing chamber re-implementation

It will be fixed in Balancing chamber re-implementation

I would prefer to fix this in balancing chamber new algorithm implementation. That is why i didn't mark it as Unresolved.

I would prefer to fix this in balancing chamber new algorithm implementation. That is why i didn't mark it as Unresolved.

fixed it

fixed it

fixed it

fixed it

This state machine running continuously. fill request is initiated for every 1 minutes of balancing chamber exec (since right now level sensor is not working ) //Increment counter for dry bicarb c...

This state machine running continuously. fill request is initiated for every 1 minutes of balancing chamber exec (since right now level sensor is not working )

//Increment counter for dry bicarb chamber fill
bicarbChamberPeriodicFillCounter += 1;
// Fill bicarb chamber once every 60secs.
if ( bicarbChamberPeriodicFillCounter >= BICARB_CHAMBER_PERIODIC_FILL_TIME )

Unknown macro: { setBicarbChamberFillRequested(); bicarbChamberPeriodicFillCounter = 0; }


The request is cleared here. for consistency I am moving to the end of the fill cycle

Mis-spelled "Stabilized".

Mis-spelled "Stabilized".

I think you can do this in 1 line of code using the RANGE macro in Common.h.

I think you can do this in 1 line of code using the RANGE macro in Common.h.

Remove blank line.

Remove blank line.

Add a blank line before return statement.

Add a blank line before return statement.

Why is this commented out?

Why is this commented out?

d48PumpSpeed is an input.

d48PumpSpeed is an input.

What should trimmer heater be doing in these spent/bicarb chamber states?

What should trimmer heater be doing in these spent/bicarb chamber states?

I would think this would be enabled by default.

I would think this would be enabled by default.