Since trimmer heater is enabled, no need of temp compensation. only for testing purposes, temp compensation is needed. Hence by default, temp compensation is disabled.

Since trimmer heater is enabled, no need of temp compensation. only for testing purposes, temp compensation is needed. Hence by default, temp compensation is disabled.

For spent chamber fill, we don't know how low the liquid level is ( in Chamber H) and hence decided not to turn on the trimmer heater for safety reasons. For bicarb fill, the dry bicart code should...

For spent chamber fill, we don't know how low the liquid level is ( in Chamber H) and hence decided not to turn on the trimmer heater for safety reasons. For bicarb fill, the dry bicart code should cover the actuators status. the current bicarb chamber fill code here should be updated with the newly implemented dry bi cart code changes.

I think cal data and count should be retrieved by same function and possibly needs some interrupt protection to ensure data and count stick together. If these functions are called from general task...

I think cal data and count should be retrieved by same function and possibly needs some interrupt protection to ensure data and count stick together. If these functions are called from general task, we could have situation where data is retrieved, then priority task updates fpga, and then counter is retrieved with wrong (incremented) counter.
If these two functions are being called from a driver that is executed in priority task, then we don't need to worry about this situation, but still ought to add a constraint/warning comment in function headers noting that these functions should only be called from priority task.

I think Dialin will need get (or UI proxy) commands to have us send cal, s/n, and versions.

I think Dialin will need get (or UI proxy) commands to have us send cal, s/n, and versions.

Do FP sensors have to be treated separately? I realize there needs to be separate messages as they will be coming from different Dialin classes, but can't we handle both messages with same function...

Do FP sensors have to be treated separately? I realize there needs to be separate messages as they will be coming from different Dialin classes, but can't we handle both messages with same function (assuming shared enum to index)?

General Dialin question - does it make sense to override temperatures at driver level where they are spread out over several drivers? I would think overriding at monitor level would be simpler wher...

General Dialin question - does it make sense to override temperatures at driver level where they are spread out over several drivers? I would think overriding at monitor level would be simpler where they are all in one place and could be indexed by enum with single command.

Do we need to say Value?

Do we need to say Value?

Add a blank line after public defs banner. Add a dox comment to the right. I believe they reserve space for 12 bytes, not 10. Not really for calibration - it's for s/n and versions.

Add a blank line after public defs banner.
Add a dox comment to the right.
I believe they reserve space for 12 bytes, not 10.
Not really for calibration - it's for s/n and versions.

Are these really going to be different types? I assumed they would all be floats.

Are these really going to be different types? I assumed they would all be floats.

Structures need a top level dox comment.

Structures need a top level dox comment.

enums need a top level dox comment.

enums need a top level dox comment.

I believe conductivity and temperature data will be F32s.

I believe conductivity and temperature data will be F32s.

Do we not need this anymore? I still see it in code all over the place and I don't think we're done w/ Teensy until Beta 1.9 units are obsolete.

Do we not need this anymore? I still see it in code all over the place and I don't think we're done w/ Teensy until Beta 1.9 units are obsolete.

Mis-spelled "Stabilized".

Mis-spelled "Stabilized".

I think you can do this in 1 line of code using the RANGE macro in Common.h.

I think you can do this in 1 line of code using the RANGE macro in Common.h.

What should trimmer heater be doing in these spent/bicarb chamber states?

What should trimmer heater be doing in these spent/bicarb chamber states?

I would think this would be enabled by default.

I would think this would be enabled by default.

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 18 more files in changeset.
resolved merge conflicts

  1. … 6 more files in changeset.
Added

Added

An alarm is triggered in this function. Add to header.

An alarm is triggered in this function. Add to header.

Updated

Updated

LEAHI-DD-FIRMWARE-LDT-2004_Dialysate Composition - DD
LEAHI-DD-FIRMWARE-LDT-2004_Dialysate Composition - DD
These values should be retained for Beta 1.0