This is a list of all comments for DG-DEN-14999-2. Review Summary: No summary ---------------------------------------- File: firmware/App/Controllers/ConductivitySensors.c Revision Comment by Michael Garthwaite on 31 March 2023, 08:52 https://devapps.diality.us/cru/DG-DEN-14999-2#c17019 will these alarms trigger with open air? Ex: if a user has a device shipped to them, will they be greeted with conductivity alarms on first power up before any water has ran through the system? If so, how will they be able to fix it? Reply by Sean Nash on 31 March 2023, 09:00 > I think VPi is closed during standby mode, so not sure why we > want to check inlet conductivity in this mode. > I imagine new system would get installed by service - would > need some kind of priming done in service mode, so I'm not > worried about that. Reply by Michael Garthwaite on 31 March 2023, 09:14 > RESOLVED IN CODE WALKTHROUGH. --- ID: DG-DEN-14999-2 https://devapps.diality.us/cru/DG-DEN-14999-2 Title: DG-DEN-14999_Alarm Table Updates Statement of Objectives: State: Closed Summary: Author: Sean Nash Moderator: Sean Nash Reviewers: (4 active, 2 completed*) Dara Navaei (*) jtaylor (*) Michael Garthwaite wbracken Darren Cox Steve Jarpe