This is a list of all comments for DG-DEN-15608-1. Review Summary: No summary ---------------------------------------- File: firmware/App/Controllers/Voltages.c Revision Comment by Sean Nash on 12 July 2023, 15:32 https://devapps.diality.us/cru/DG-DEN-15608-1#c18084 Won't this voltage go to zero when heater is off? Reply by wbracken on 12 July 2023, 15:48 > It is only monitored when not in safe state. Reviewed with > Blaine. Reply by Sean Nash on 12 July 2023, 15:51 > But you can be "not in safe state" and not have the heater > on (e.g. in standby mode). Is this voltage always 24V even > when the heater is off? Reply by Sean Nash on 12 July 2023, 15:54 > RESOLVED in CODE WALKTHROUGH. --- ID: DG-DEN-15608-1 https://devapps.diality.us/cru/DG-DEN-15608-1 Title: DG-DEN-15608_Alarm 77 Failing TO Trigger For Lower Bound Values Statement of Objectives: State: Closed Summary: Author: wbracken Moderator: wbracken Reviewers: (4 active, 2 completed*) Sean Nash (*) jtaylor (*) Michael Garthwaite Dara Navaei Darren Cox Steve Jarpe