This is a list of all comments for DG-DEN-15842-1. Review Summary: No summary ---------------------------------------- File: AlarmDefs.h Revision Comment by Sean Nash on 20 July 2023, 15:34 https://devapps.diality.us/cru/DG-DEN-15842-1#c18217 Are we sure we want this? Reply by Darren Cox on 20 July 2023, 16:04 > Blaine's comment shows discrepancy between HD and DG fan > handling with these flags. Would we ever want to block an HD > Fan alarm? Looks to indicate a HW fail. Reply by Sean Nash on 20 July 2023, 17:18 > It's not a fault though. User can choose to resume, > rinseback or end tx but will the alarm just trigger again > immediately if you haven't resolved the h/w problem? With > the 2 properties set to TRUE, user can choose to rinseback > or end tx and they will not be bothered by this alarm again > until rinseback and post-tx are completed. Reply by Darren Cox on 27 July 2023, 14:02 > I spoke with Blaine, he said make them consistent and > defer to Sean. So they will both be as HD was - No > re-trigger in Rinseback or End TX. Reply by Dara Navaei on 27 July 2023, 14:57 > RESOLVED in CODE WALKTHROUGH. --- ID: DG-DEN-15842-1 https://devapps.diality.us/cru/DG-DEN-15842-1 Title: DG-DEN-15842_Alarm 272 Fan Fault HD Alarm Does Not Declare During Post TX Statement of Objectives: State: Closed Summary: Author: Darren Cox Moderator: Darren Cox Reviewers: (4 active, 2 completed*) Dara Navaei (*) jtaylor (*) Sean Nash Michael Garthwaite wbracken Steve Jarpe