This is a list of all comments for DG-DENBUG-307-1. Review Summary: No summary ---------------------------------------- File: firmware/App/Controllers/ConcentratePumps.c Revision Comment by Sean Nash on 01 April 2025, 08:59 https://devapps.diality.us/cru/DG-DENBUG-307-1#c21859 Measured speed (whether real or override) should always be positive. And if that's true, we shouldn't need this exception. Reply by Dara Navaei on 02 April 2025, 10:28 > Done Reply by Sean Nash on 02 April 2025, 11:13 > RESOLVED in CODE WALKTHROUGH. Revision Comment by Sean Nash on 01 April 2025, 09:01 https://devapps.diality.us/cru/DG-DENBUG-307-1#c21860 Aren't speeds (commanded and measured) always positive? Do we need the fabs? Reply by Dara Navaei on 02 April 2025, 10:26 > I used fabs so all are positive when monitoring is started. Reply by Sean Nash on 02 April 2025, 11:13 > Not needed, but benign. RESOLVED in CODE WALKTHROUGH. Revision Comment by Sean Nash on 01 April 2025, 09:03 https://devapps.diality.us/cru/DG-DENBUG-307-1#c21861 This check doesn't make sense to me. Measured speed will only be negative if set direction is reverse. So how could directions ever not match? Reply by Dara Navaei on 02 April 2025, 10:27 > Done Reply by Sean Nash on 02 April 2025, 11:13 > RESOLVED in CODE WALKTHROUGH. --- ID: DG-DENBUG-307-1 https://devapps.diality.us/cru/DG-DENBUG-307-1 Title: DG-DENBUG-307_Concentrate Pump Tube Ruptures During Heat Disinfection Staging Statement of Objectives: State: Closed Summary: Author: Dara Navaei Moderator: Dara Navaei Reviewers: (4 active, 2 completed*) Sean Nash (*) Michael Garthwaite (*) jpaguio Vinayakam Mani amanesh Daniel Ho