This is a list of all comments for DIALIN-DEN-3433-1. Review Summary: No summary ---------------------------------------- File: tests/test_alarms.py Revision Comment by pmontazemi on 11 June 2020, 16:07 https://devapps.diality.us/cru/DIALIN-DEN-3433-1#c2263 Are these the only alarm faults for the entire Denali machine? Reply by plucia on 11 June 2020, 16:13 > No these are Non-Fault alarms that are recoverable. > > I've excluded Fault alarms in this specific script since that > would require a manual reset of the HD between each > consecutive alarm test. > > All possible alarms can be found in alarms.py Reply by pmontazemi on 11 June 2020, 16:18 > RESOLVED. --- ID: DIALIN-DEN-3433-1 https://devapps.diality.us/cru/DIALIN-DEN-3433-1 Title: DIALIN-DEN-3433_Maintenance Statement of Objectives: State: Closed Summary: Author: plucia Moderator: plucia Reviewers: (0 active, 2 completed*) Sean Nash (*) pmontazemi (*)