This is a list of all comments for HD-DEN-14011-1. Review Summary: No summary ---------------------------------------- File: firmware/App/Controllers/PresOccl.c Revision Comment by Sean Nash on 20 January 2023, 09:06 https://devapps.diality.us/cru/HD-DEN-14011-1#c16172 Not sure exception for treatment stop state still makes sense. We can be running BP in that state now depending on alarm. Should still alarm on occlusion. Reply by Michael Garthwaite on 20 January 2023, 11:27 > Fixed. Only checking if BP is running when are in treatment > mode. Reply by Sean Nash on 20 January 2023, 11:38 > RESOLVED in CODE WALKTHROUGH. ---------------------------------------- File: firmware/App/Modes/SelfTests.c Revision Comment by Sean Nash on 20 January 2023, 13:59 https://devapps.diality.us/cru/HD-DEN-14011-1#c16189 Should set pump track closed required, but not door here. Reply by Michael Garthwaite on 20 January 2023, 15:54 > Fixed. Thanks! Reply by Sean Nash on 20 January 2023, 16:01 > RESOLVED in CODE WALKTHROUGH. Revision Comment by Sean Nash on 20 January 2023, 09:11 https://devapps.diality.us/cru/HD-DEN-14011-1#c16173 Should this 10 seconds only start once latch is locked? Reply by Michael Garthwaite on 20 January 2023, 11:28 > Fixed. Created handleDrySelfTestStartState() that counts > settle time once the latch is determined closed. Reply by Sean Nash on 20 January 2023, 11:35 > RESOLVED in CODE WALKTHROUGH. Revision Comment by Sean Nash on 20 January 2023, 14:00 https://devapps.diality.us/cru/HD-DEN-14011-1#c16191 transition and resume part of stopped state should set pump track required so we don't have to keep setting it here every 50 ms. Reply by Michael Garthwaite on 20 January 2023, 15:54 > Fixed. Thanks! Reply by Sean Nash on 20 January 2023, 16:00 > RESOLVED in CODE WALKTHROUGH. Revision Comment by Sean Nash on 20 January 2023, 14:00 https://devapps.diality.us/cru/HD-DEN-14011-1#c16190 Should set door and pump track required here before we move on to next state. Reply by Michael Garthwaite on 20 January 2023, 15:54 > Fixed. Thanks! Reply by Sean Nash on 20 January 2023, 16:01 > RESOLVED in CODE WALKTHROUGH. Revision Comment by Sean Nash on 20 January 2023, 11:36 https://devapps.diality.us/cru/HD-DEN-14011-1#c16182 Do we need an else where we reset 10 second start time (if latch opened)? Reply by Sean Nash on 20 January 2023, 14:02 > RESOLVED in CODE WALKTHROUGH. Revision Comment by Sean Nash on 20 January 2023, 09:13 https://devapps.diality.us/cru/HD-DEN-14011-1#c16174 Should this be moved into "if" body? Doesn't appear to need setting until we resume. Reply by Michael Garthwaite on 20 January 2023, 11:28 > Fixed. Thanks! Reply by Sean Nash on 20 January 2023, 11:34 > RESOLVED in CODE WALKTHROUGH. Revision Comment by Sean Nash on 20 January 2023, 14:05 https://devapps.diality.us/cru/HD-DEN-14011-1#c16193 If going back to start state, set pump track closed required, but not door closed. If skipping dry self-tests, set both to be required closed. Also, this s/w config looks like it's setup wrong. For release build, we want to go back to start state. Reply by Michael Garthwaite on 20 January 2023, 15:54 > Fixed. Thanks! Reply by Sean Nash on 20 January 2023, 15:59 > RESOLVED in CODE WALKTHROUGH. --- ID: HD-DEN-14011-1 https://devapps.diality.us/cru/HD-DEN-14011-1 Title: HD-DEN-14011_Occulsion Alarm When Reseating Statement of Objectives: State: Closed Summary: Author: Michael Garthwaite Moderator: Michael Garthwaite Reviewers: (4 active, 2 completed*) Sean Nash (*) wbracken (*) Dara Navaei Darren Cox jtaylor Steve Jarpe