This is a list of all comments for HD-DEN-14252-1. Review Summary: No summary ---------------------------------------- File: firmware/App/Modes/Prime.c Revision Comment by jtaylor on 02 December 2022, 09:12 https://devapps.diality.us/cru/HD-DEN-14252-1#c15094 It seems strange to add a magic number here, from a defined constant. I'd assume (guess) the relation between 300->225 is maintaining a delta P along the flow path (or across the membrane?), but it might be worth spelling out, at the top with the defines, or as a macro function of the input flow rate. Reply by Sean Nash on 02 December 2022, 16:37 > Created a #define for magic number. > Because running pumps in open loop mode (because we're > priming and not necessarily any fluid in line to measure flow > with) and pumps have different methods of determining PWM > duty cycle from a given flow rate, DPo pumps target flow rate > needed to be set lower to match the duty cycle with DPi pump. Reply by jtaylor on 05 December 2022, 13:42 > RESOLVED in CODE WALKTHROUGH. --- ID: HD-DEN-14252-1 https://devapps.diality.us/cru/HD-DEN-14252-1 Title: HD-DEN-14252_System Self Test Failure Occurs Intermittently Statement of Objectives: State: Closed Summary: Author: Dara Navaei Moderator: Dara Navaei Reviewers: (2 active, 2 completed*) Sean Nash (*) jtaylor (*) wbracken Michael Garthwaite