This is a list of all comments for HD-DEN-14546-3. Review Summary: No summary ---------------------------------------- File: firmware/App/Controllers/SyringePump.c Revision Comment by Sean Nash on 24 April 2023, 10:31 https://devapps.diality.us/cru/HD-DEN-14546-3#c17226 Why accept Heparin resume when our syringe is empty? Reply by jtaylor on 27 April 2023, 15:57 > The suggested mechanism to resume treatment without heparin > had been to allow resume with minor flag changes to allow the > treatment to continue. At this point clearing the alarm may > be all that is necessary. Code updated. --- ID: HD-DEN-14546-3 https://devapps.diality.us/cru/HD-DEN-14546-3 Title: HD-DEN-14546_Add Option TO Resume Treatment Without Heparin For Empty Syringe Alarm Statement of Objectives: State: Closed Summary: Author: jtaylor Moderator: jtaylor Reviewers: (6 active, 0 completed*) Sean Nash Michael Garthwaite wbracken Dara Navaei Darren Cox Steve Jarpe