This is a list of all comments for HD-DEN-14830-2. Review Summary: No summary ---------------------------------------- File: firmware/App/Controllers/DGInterface.c Revision Comment by Sean Nash on 27 February 2023, 11:29 https://devapps.diality.us/cru/HD-DEN-14830-2#c16532 I thought the non-safety temp alarms had a large persistence. Now they have none? Reply by Michael Garthwaite on 27 February 2023, 11:35 > PRS (Ex: PRS 377) no longer has it as of 1/05. SRS (Ex: SRSHD > 901) does. which one should be enforced? Reply by Sean Nash on 27 February 2023, 11:39 > Let's review PRS req together. Seems like we have > conflicting alarm properties in place that don't make sense > to me. Let's figure out what behavior Systems team is > looking for here and then decide what > properties/persistence should be set. Reply by Michael Garthwaite on 07 March 2023, 16:42 > Fixed. Thanks! Reply by Sean Nash on 08 March 2023, 21:12 > RESOLVED in CODE WALKTHROUGH. Revision Comment by Sean Nash on 08 March 2023, 21:05 https://devapps.diality.us/cru/HD-DEN-14830-2#c16659 Remove blank line. Reply by Sean Nash on 07 April 2023, 13:29 > RESOLVED in CODE WALKTHROUGH. ---------------------------------------- File: AlarmDefs.h Revision Comment by wbracken on 05 April 2023, 13:44 https://devapps.diality.us/cru/HD-DEN-14830-2#c17049 Comment alignment. Reply by Michael Garthwaite on 07 April 2023, 11:48 > Fixed. Thanks! Reply by wbracken on 07 April 2023, 13:32 > RESOLVED IN CODE WALKTHROUGH Revision Comment by Sean Nash on 27 February 2023, 11:22 https://devapps.diality.us/cru/HD-DEN-14830-2#c16531 If you clear immediately, it will let you resume/clear alarm immediately (before temp comes back into range), but then alarm will re-trigger immediately (because temp is still out of range). Reply by Michael Garthwaite on 07 March 2023, 16:54 > Fixed? Reply by Sean Nash on 08 March 2023, 21:15 > RESOLVED in CODE WALKTHROUGH. Revision Comment by Sean Nash on 06 March 2023, 09:58 https://devapps.diality.us/cru/HD-DEN-14830-2#c16586 Why alarm if above low temp? I think >42 or > tgt+4 should be two ways to get same non-safety high temp alarm. And I think <33 or < tgt-4 should be two ways to get same non-safety low temp alarm. Did Systems want to separate these? My understanding was not to separate. Reply by Michael Garthwaite on 07 March 2023, 16:46 > To answer this and comment below: > > above_high_safety_range = temp > 46 > above_low_safety_range = temp > 42 > above_target_temp = target + 4 > below_target_temp = target - 4 > below_low_safety_range = temp < 33 > > I did not see any specification regarding to separate or not > to separate. Original functionality had separate triggers. > Respective PRS's state OR between the two triggers. > I think it would be clear to for us in terms of debugging to > see which alarm is triggered since on the low side, temp - 4 > would be lower than 33, (31). > > Wording may be unclear. Im up for suggestions to help clarify > the alarm differences. Reply by Sean Nash on 07 April 2023, 13:30 > RESOLVED in CODE WALKTHROUGH. --- ID: HD-DEN-14830-2 https://devapps.diality.us/cru/HD-DEN-14830-2 Title: HD-DEN-14830_Incorrect Dialysate Temperature Alarm Triggered Statement of Objectives: State: Closed Summary: Author: Michael Garthwaite Moderator: Michael Garthwaite Reviewers: (3 active, 3 completed*) Sean Nash (*) wbracken (*) Dara Navaei (*) Darren Cox jtaylor Steve Jarpe