This is a list of all comments for HD-DEN-15049-1. Review Summary: No summary ---------------------------------------- File: firmware/App/Controllers/Voltages.c Revision Comment by Sean Nash on 13 April 2023, 13:14 https://devapps.diality.us/cru/HD-DEN-15049-1#c17092 No longer needed? Reply by wbracken on 13 April 2023, 13:48 > Removed Reply by Sean Nash on 13 April 2023, 13:58 > RESOLVED in CODE WALKTHROUGH. Revision Comment by Sean Nash on 13 April 2023, 13:10 https://devapps.diality.us/cru/HD-DEN-15049-1#c17088 We want to use CPLD for this. Keep this line, remove line below where initialized to FALSE. Reply by wbracken on 13 April 2023, 13:53 > Corrected Reply by Sean Nash on 13 April 2023, 13:58 > RESOLVED in CODE WALKTHROUGH. Revision Comment by Sean Nash on 13 April 2023, 13:12 https://devapps.diality.us/cru/HD-DEN-15049-1#c17090 Remove this method of determine whether AC power has been lost. CPLD function above is what we want. Reply by wbracken on 13 April 2023, 13:52 > Corrected Reply by Sean Nash on 13 April 2023, 13:57 > RESOLVED in CODE WALKTHROUGH. Revision Comment by Sean Nash on 13 April 2023, 13:16 https://devapps.diality.us/cru/HD-DEN-15049-1#c17093 I think we can fit this whole comment on the 1 line. Bring closer to break; if need more room. Reply by wbracken on 13 April 2023, 13:51 > Corrected Reply by Sean Nash on 13 April 2023, 13:57 > RESOLVED in CODE WALKTHROUGH. Revision Comment by Sean Nash on 13 April 2023, 13:20 https://devapps.diality.us/cru/HD-DEN-15049-1#c17094 I think we can just say "if ( TRUE == hasPowerBeenLost )" here. Zero trigger persistence time now that CPLD is making determination. Reply by wbracken on 13 April 2023, 13:51 > Corrected Reply by Sean Nash on 13 April 2023, 13:57 > RESOLVED in CODE WALKTHROUGH. Revision Comment by Darren Cox on 13 April 2023, 14:17 https://devapps.diality.us/cru/HD-DEN-15049-1#c17112 Are these alarms intentionally commented? Should they be activated or removed? Reply by wbracken on 13 April 2023, 14:22 > They should remain. Will be uncommented when all DVT units > have updated hardware. Reply by Dara Navaei on 19 October 2023, 09:04 > RESOLVED in CODE WALKTHROUGH ---------------------------------------- File: firmware/App/Controllers/Voltages.h Revision Comment by Sean Nash on 12 April 2023, 11:05 https://devapps.diality.us/cru/HD-DEN-15049-1#c17077 Comments appear to be swapped. Reply by wbracken on 12 April 2023, 11:24 > Comment is identical to the DG implementation. Reply by Sean Nash on 13 April 2023, 11:46 > Still wrong though. Reply by wbracken on 13 April 2023, 13:11 > Corrected Reply by Sean Nash on 13 April 2023, 13:13 > RESOLVED in CODE WALKTHROUGH. --- ID: HD-DEN-15049-1 https://devapps.diality.us/cru/HD-DEN-15049-1 Title: HD-DEN-15049_FW HD Voltage Not Trigger Alarm 76 When Voltage Out OF Range IN Line Statement of Objectives: State: Closed Summary: Author: wbracken Moderator: wbracken Reviewers: (4 active, 2 completed*) Sean Nash (*) jtaylor (*) Michael Garthwaite Dara Navaei Darren Cox Steve Jarpe