This is a list of all comments for HD-DEN-15716-1. Review Summary: No summary ---------------------------------------- File: firmware/App/Modes/SelfTests.c Revision Comment by Sean Nash on 11 July 2023, 12:34 https://devapps.diality.us/cru/HD-DEN-15716-1#c18039 The 2 verify states are single pass. I don't think it makes sense to return to them. They either pass or you go to fault mode and so you wouldn't even be here - so I think you can assume you passed. For first displacement verify state, I think you should go to second displacement setup state. For second displacement verify state, I think you should go to complete state. Reply by Michael Garthwaite on 11 July 2023, 13:24 > Fixed. Thanks! Reply by Sean Nash on 12 July 2023, 15:43 > RESOLVED in CODE WALKTHROUGH. --- ID: HD-DEN-15716-1 https://devapps.diality.us/cru/HD-DEN-15716-1 Title: HD-DEN-15716_Skipped Treatment Start State While Transitioning From Pre Treatment TO Treatment Statement of Objectives: State: Closed Summary: Author: Michael Garthwaite Moderator: Michael Garthwaite Reviewers: (4 active, 2 completed*) wbracken (*) Dara Navaei (*) Sean Nash Darren Cox jtaylor Steve Jarpe