This is a list of all comments for HD-DEN-16391-3. Review Summary: No summary ---------------------------------------- File: firmware/App/Controllers/DGInterface.c Revision Comment by Dara Navaei on 02 January 2024, 11:06 https://devapps.diality.us/cru/HD-DEN-16391-3#c19329 Why was the = removed from the ternary ifs? Reply by Vinayakam Mani on 02 January 2024, 14:10 > Discussed with Sean and confirmed that matching requirement > done in the code. We are good with the modification now. Reply by Dara Navaei on 03 January 2024, 16:22 > RESOLVED in CODE WALKTHROUGH --- ID: HD-DEN-16391-3 https://devapps.diality.us/cru/HD-DEN-16391-3 Title: HD-DEN-16391_FW Dialysate Temperature Alarm Not Sensitive Enough Use Tdi Moving Avg Statement of Objectives: State: Closed Summary: Author: Vinayakam Mani Moderator: Vinayakam Mani Reviewers: (3 active, 2 completed*) Sean Nash (*) Dara Navaei (*) jpaguio Michael Garthwaite Darren Cox