This is a list of all comments for HD-DEN-16512-1. Review Summary: No summary ---------------------------------------- File: firmware/App/Controllers/DialOutFlow.c Revision Comment by Sean Nash on 27 November 2023, 09:26 https://devapps.diality.us/cru/HD-DEN-16512-1#c19264 5% seems like a reasonable max step. I don't think we want to go to ramp up/down states if > 5% change. I think we just want to limit the change to max 5% for a single control interval. Reply by Vinayakam Mani on 27 November 2023, 13:34 > Done. Reply by Sean Nash on 28 November 2023, 09:24 > RESOLVED in CODE WALKTHROUGH. Revision Comment by Sean Nash on 27 November 2023, 09:24 https://devapps.diality.us/cru/HD-DEN-16512-1#c19262 Add space after if. Reply by Vinayakam Mani on 27 November 2023, 13:35 > Done. Reply by Sean Nash on 28 November 2023, 09:23 > RESOLVED in CODE WALKTHROUGH. Revision Comment by Sean Nash on 27 November 2023, 09:25 https://devapps.diality.us/cru/HD-DEN-16512-1#c19263 Remove blank line. Reply by Vinayakam Mani on 27 November 2023, 13:34 > Done. Reply by Sean Nash on 28 November 2023, 09:25 > RESOLVED in CODE WALKTHROUGH. ---------------------------------------- File: firmware/App/Modes/ModeTreatment.c Revision Comment by Sean Nash on 04 December 2023, 14:22 https://devapps.diality.us/cru/HD-DEN-16512-1#c19302 Consider splitting this if and the setDialysisParams() function into two: 1) blood flow rate changed and 2) dialysate flow rate changed. Reply by Vinayakam Mani on 06 December 2023, 16:05 > Done. Reply by Sean Nash on 07 December 2023, 09:41 > RESOLVED in CODE WALKTHROUGH. --- ID: HD-DEN-16512-1 https://devapps.diality.us/cru/HD-DEN-16512-1 Title: HD-DEN-16512_Dpo Stops During Treatment After Reconfirming Dialysate Flow Rate Statement of Objectives: State: Closed Summary: Author: Vinayakam Mani Moderator: Vinayakam Mani Reviewers: (5 active, 2 completed*) Sean Nash (*) Dara Navaei (*) jpaguio Michael Garthwaite wbracken Darren Cox jtaylor