This is a list of all comments for HD-DEN-16595-2. Review Summary: No summary ---------------------------------------- File: firmware/App/Services/AlarmMgmt.c Revision Comment by Dara Navaei on 08 January 2024, 08:56 https://devapps.diality.us/cru/HD-DEN-16595-2#c19342 Why was this block removed? We do not need to check of the empty bottles alarms anymore? Reply by Vinayakam Mani on 08 January 2024, 12:59 > The removed code changes are dead code due to condition of > "if" condition check at top level ( ALARM_TABLE[ a > ].alarmClearOnly). these alarms have property of 'clear only' > as 'TRUE" in the alarm table. hence condition always be false > and the removed code changes never executed. Reply by Dara Navaei on 08 January 2024, 20:10 > RESOLVED in CODE WALKTHROUGH --- ID: HD-DEN-16595-2 https://devapps.diality.us/cru/HD-DEN-16595-2 Title: HD-DEN-16595_Vectorcast Unit And Integration Testing Statement of Objectives: State: Closed Summary: Author: Vinayakam Mani Moderator: Vinayakam Mani Reviewers: (3 active, 2 completed*) Sean Nash (*) Dara Navaei (*) jpaguio Michael Garthwaite Darren Cox