This is a list of all comments for TD-LEAH-51-1. Review Summary: No summary ---------------------------------------- File: firmware/App/Controllers/BloodFlow.c Revision Comment by Vinayakam Mani on 01 November 2024, 15:59 https://devapps.diality.us/cru/TD-LEAH-51-1#c20735 There is another macro " MAX_SETTABLE_BLOOD_FLOW_RATE" defining 700.. are they similar or different one? may need to change macro name or add comment for clarification. Reply by Sean Nash on 04 November 2024, 11:53 > Per comment, 1320 is maximum "measured" flow rate. 700 is > maximum "settable" flow rate. 500 is maximum prescribed flow > rate. Revision Comment by Vinayakam Mani on 01 November 2024, 16:23 https://devapps.diality.us/cru/TD-LEAH-51-1#c20736 Please add \b style formatting for inputs, outputs and alarms (entire file). Also add alarm details where needed. ---------------------------------------- File: firmware/App/Controllers/BloodFlow.h Revision Comment by Vinayakam Mani on 01 November 2024, 16:27 https://devapps.diality.us/cru/TD-LEAH-51-1#c20737 Please comment comparing the previous macros in .c file. ---------------------------------------- File: firmware/App/Controllers/Valves.c Revision Comment by Vinayakam Mani on 25 October 2024, 16:25 https://devapps.diality.us/cru/TD-LEAH-51-1#c20614 Valve ( V caps)? Reply by Sean Nash on 28 October 2024, 09:37 > Fixed. Revision Comment by Vinayakam Mani on 25 October 2024, 16:50 https://devapps.diality.us/cru/TD-LEAH-51-1#c20615 Please align the comments to fit in same column, comparing the top comments. Reply by Sean Nash on 28 October 2024, 09:37 > Fixed. Revision Comment by Vinayakam Mani on 25 October 2024, 17:01 https://devapps.diality.us/cru/TD-LEAH-51-1#c20616 Valves broadcast Reply by Sean Nash on 28 October 2024, 09:37 > Fixed. ---------------------------------------- File: firmware/App/Drivers/LevelSensors.c Revision Comment by Vinayakam Mani on 28 October 2024, 09:26 https://devapps.diality.us/cru/TD-LEAH-51-1#c20621 Do we need this timeout to be changed as task interval counter as we were seeing additional 100ms debounce time at times (hence adjusted the requirement)? Reply by Sean Nash on 28 October 2024, 10:07 > Yes. Changed to task interval based debounce timing. ---------------------------------------- File: firmware/App/Drivers/PeristalticPump.c Revision Comment by Vinayakam Mani on 28 October 2024, 09:42 https://devapps.diality.us/cru/TD-LEAH-51-1#c20627 unit variables initialized. Reply by Sean Nash on 28 October 2024, 10:10 > Fixed. Revision Comment by Vinayakam Mani on 28 October 2024, 09:45 https://devapps.diality.us/cru/TD-LEAH-51-1#c20628 looks the same function name used in FpgaTD and BloodFlow.C files. Do we want to differentiate between these two-function naming? Reply by Sean Nash on 28 October 2024, 10:15 > Fixed. Revision Comment by Vinayakam Mani on 28 October 2024, 09:52 https://devapps.diality.us/cru/TD-LEAH-51-1#c20629 Message payload length check to be considered. Reply by Sean Nash on 28 October 2024, 10:24 > Fixed. ---------------------------------------- File: firmware/App/Drivers/RotaryValve.c Revision Comment by Vinayakam Mani on 28 October 2024, 10:10 https://devapps.diality.us/cru/TD-LEAH-51-1#c20631 Please align the comments into same column. Reply by Sean Nash on 29 October 2024, 09:07 > Fixed. Revision Comment by Vinayakam Mani on 28 October 2024, 10:20 https://devapps.diality.us/cru/TD-LEAH-51-1#c20634 payload length check to be considered. Reply by Sean Nash on 28 October 2024, 10:24 > Fixed. ---------------------------------------- File: firmware/App/Drivers/Valve2Way.c Revision Comment by Vinayakam Mani on 28 October 2024, 10:23 https://devapps.diality.us/cru/TD-LEAH-51-1#c20635 unit variables initialized. Reply by Sean Nash on 29 October 2024, 09:05 > Fixed. ---------------------------------------- File: firmware/App/Drivers/Valve2Way.h Revision Comment by Vinayakam Mani on 28 October 2024, 10:26 https://devapps.diality.us/cru/TD-LEAH-51-1#c20638 Remove 'a'. Reply by Sean Nash on 28 October 2024, 10:27 > Fixed. ---------------------------------------- File: firmware/App/Services/AlarmMgmtTD.c Revision Comment by Vinayakam Mani on 28 October 2024, 10:31 https://devapps.diality.us/cru/TD-LEAH-51-1#c20640 payload length check. Reply by Sean Nash on 29 October 2024, 09:04 > Fixed. --- ID: TD-LEAH-51-1 https://devapps.diality.us/cru/TD-LEAH-51-1 Title: TD-LEAH-51_FW TD Blood Pump Controller Moni Statement of Objectives: State: Closed Summary: Author: Sean Nash Moderator: Sean Nash Reviewers: (2 active, 2 completed*) Vinayakam Mani (*) Michael Garthwaite (*) jpaguio Dara Navaei