This is a list of all comments for TD-LEAH-838-1. Review Summary: No summary ---------------------------------------- File: firmware/App/Controllers/AirPump.c Revision Comment by Vinayakam Mani on 20 February 2025, 14:07 https://devapps.diality.us/cru/TD-LEAH-838-1#c21506 Please add in outputs: currentAirPumpMotorPowerLevel Reply by Sean Nash on 21 February 2025, 14:00 > Done. ---------------------------------------- File: firmware/App/Controllers/AirTrap.c Revision Comment by Vinayakam Mani on 20 February 2025, 15:00 https://devapps.diality.us/cru/TD-LEAH-838-1#c21511 I remember there was a change done in Denali to extend the timeout value. Do we need to adjust the timeout value for LEAHI? Reply by Sean Nash on 21 February 2025, 14:09 > Change has not happened yet in Denali. I will follow up with > Dara after change is made. Revision Comment by Vinayakam Mani on 20 February 2025, 14:47 https://devapps.diality.us/cru/TD-LEAH-838-1#c21507 Do we want to remove the initialization (FALSE), as it is getting initialized as part of the init process. Reply by Sean Nash on 21 February 2025, 14:10 > Yes. Done. Revision Comment by Vinayakam Mani on 20 February 2025, 14:51 https://devapps.diality.us/cru/TD-LEAH-838-1#c21508 Remove events message. Reply by Sean Nash on 21 February 2025, 14:11 > Done. Revision Comment by Vinayakam Mani on 20 February 2025, 14:55 https://devapps.diality.us/cru/TD-LEAH-838-1#c21509 Do we need to rename LevelSensors to AirTrapLevelSensors to get right context? Reply by Sean Nash on 21 February 2025, 14:13 > Done. Revision Comment by Vinayakam Mani on 20 February 2025, 14:58 https://devapps.diality.us/cru/TD-LEAH-838-1#c21510 Remove additional space Reply by Sean Nash on 21 February 2025, 14:14 > Done. Revision Comment by Vinayakam Mani on 20 February 2025, 15:20 https://devapps.diality.us/cru/TD-LEAH-838-1#c21513 #define the 1 and 0. Reply by Sean Nash on 21 February 2025, 14:23 > Done. Revision Comment by Vinayakam Mani on 20 February 2025, 15:15 https://devapps.diality.us/cru/TD-LEAH-838-1#c21512 I guess air trap monitor function handles the timeout function and no need of handling time out here (TODO). Reply by Sean Nash on 21 February 2025, 14:25 > Removed TODO. Revision Comment by Vinayakam Mani on 20 February 2025, 15:31 https://devapps.diality.us/cru/TD-LEAH-838-1#c21515 please add description/details along with the Message ID Reply by Sean Nash on 21 February 2025, 14:30 > Done. Revision Comment by Vinayakam Mani on 20 February 2025, 15:30 https://devapps.diality.us/cru/TD-LEAH-838-1#c21514 Do we need explicit typecast to U32? Reply by Sean Nash on 21 February 2025, 14:27 > I would say we are required to explicitly typecast if any of > the following are true: > 1) compiler warning if you don't > 2) byte count/mem size matters (e.g. message payload) and > might not match desired size if you don't cast it > > I don't think either is true in this case. ---------------------------------------- File: firmware/App/Controllers/AirTrap.h Revision Comment by Vinayakam Mani on 20 February 2025, 15:33 https://devapps.diality.us/cru/TD-LEAH-838-1#c21516 isAutoControlling? Reply by Sean Nash on 21 February 2025, 14:32 > Done. ---------------------------------------- File: firmware/App/Drivers/GLXferPump.c Revision Comment by Vinayakam Mani on 20 February 2025, 15:46 https://devapps.diality.us/cru/TD-LEAH-838-1#c21518 MotorState to MotorPowerLevel - across file. Reply by Sean Nash on 21 February 2025, 14:36 > Done. Revision Comment by Vinayakam Mani on 20 February 2025, 15:49 https://devapps.diality.us/cru/TD-LEAH-838-1#c21519 No Alarm, Add Event ID. Reply by Sean Nash on 21 February 2025, 14:38 > Done. Revision Comment by Vinayakam Mani on 20 February 2025, 15:51 https://devapps.diality.us/cru/TD-LEAH-838-1#c21520 May be add a check to validate the power level between 0 to 255. Reply by Sean Nash on 21 February 2025, 14:42 > Done. ---------------------------------------- File: firmware/App/Drivers/Valve3Way.c Revision Comment by Vinayakam Mani on 20 February 2025, 15:57 https://devapps.diality.us/cru/TD-LEAH-838-1#c21521 3-way Reply by Sean Nash on 21 February 2025, 14:42 > Fixed. Revision Comment by Vinayakam Mani on 20 February 2025, 16:01 https://devapps.diality.us/cru/TD-LEAH-838-1#c21522 Do we want to change STATE_CLOSED to CMN_TO_NO_OPEN (or similar?), STATE_OPEN to CMN _TO_NC_OPEN? Reply by Sean Nash on 21 February 2025, 15:02 > Done. ---------------------------------------- File: firmware/App/Controllers/BloodFlow.c Revision Comment by Vinayakam Mani on 20 February 2025, 15:38 https://devapps.diality.us/cru/TD-LEAH-838-1#c21517 Please adjust the alignment(few #defines) Reply by Sean Nash on 21 February 2025, 14:34 > Done. --- ID: TD-LEAH-838-1 https://devapps.diality.us/cru/TD-LEAH-838-1 Title: TD-LEAH-838_FW TD Gas Liquid Transfer Pump Statement of Objectives: State: Closed Summary: Author: Sean Nash Moderator: Sean Nash Reviewers: (2 active, 2 completed*) Vinayakam Mani (*) Michael Garthwaite (*) jpaguio Dara Navaei