Objectives
- DENBUG-136: Main treatment screen fails to prompt user with vital measurements entry after set interval
- the issue cannot be reproduced.
- added a count down timer for the vital entry to better understand how and when the timer works on each interval.
- DENBUG-179, DENBUG-79: Export dialog files count issues.
DENBUG-179: Export Logs text does not update after Factory Reset
DENBUG-79: UI: SD-Card file count increments when scrolling up and down the log file window
OPEN - DEN-12728: UI: UI does not clean up Treatment Parameter settings for different patient
- the reset is added to the treatment parameters
- the foundation of the pre-population has been added but still there is parts need to be added to store the sliders' On/Off switch to be restored which is needed for the slider to set properly.
- DENBUG-182: Phantom Alarm Screen Appears After Clearing Last Alarm
- the order of the QML PROPERTYs cannot exactly predicted therefore setting them if not needed has to be avoided.
- set the alarmID to 0 to trigger the next change since we skip the property updated if the alarmID is 0.