HD-DEN-14099

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Cleaned up missed artifacts from moving the bubble pressure code.

Cleaned up missed artifacts from moving the bubble pressure code.

This state should no longer be pressurizing (moved to blood prime state). Remove opening of VDi valve - no longer needed or appropriate.

This state should no longer be pressurizing (moved to blood prime state). Remove opening of VDi valve - no longer needed or appropriate.

Code is removed. RESOLVED in CODE WALKTHROUGH.

Code is removed. RESOLVED in CODE WALKTHROUGH.

RESOLVED IN CODE WALKTHROUGH.

RESOLVED IN CODE WALKTHROUGH.

Updated to reset valve position and flag.

Updated to reset valve position and flag.

Added define. There is no harm in closing valve if already closed. TODO comment below is a place holder for checking if closed too long. More testing by systems needed to determine if this is neces...

Added define. There is no harm in closing valve if already closed. TODO comment below is a place holder for checking if closed too long. More testing by systems needed to determine if this is necessary.

Updated to venous. Good catch.

Updated to venous. Good catch.

please fix naming. comment says arterial but define is calling it as venous.

please fix naming. comment says arterial but define is calling it as venous.

If this condition (timeout) occurs in middle of pressure build up, will we release on transition to next state?

If this condition (timeout) occurs in middle of pressure build up, will we release on transition to next state?

Magic # 5 - use #define. What happens if prior bubble clear is still active when this times out? Seems like condition should check that bubbleClearActive is FALSE too.

Magic # 5 - use #define.
What happens if prior bubble clear is still active when this times out? Seems like condition should check that bubbleClearActive is FALSE too.

HD-DEN-14099_Bubbles IN Dialysate Out Line From DG TO Cartridge
HD-DEN-14099_Bubbles IN Dialysate Out Line From DG TO Cartridge