HD-DEN-14830

HD-DEN-14830_Incorrect Dialysate Temperature Pt2
HD-DEN-14830_Incorrect Dialysate Temperature Pt2
RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Fixed. Thanks!

Fixed. Thanks!

Comment alignment.

Comment alignment.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Remove blank line.

Remove blank line.

Fixed?

Fixed?

To answer this and comment below: above_high_safety_range = temp > 46 above_low_safety_range = temp > 42 above_target_temp = target + 4 below_target_temp = target - 4 below_low_safety_range = temp...

To answer this and comment below:

above_high_safety_range = temp > 46
above_low_safety_range = temp > 42
above_target_temp = target + 4
below_target_temp = target - 4
below_low_safety_range = temp < 33

I did not see any specification regarding to separate or not to separate. Original functionality had separate triggers. Respective PRS's state OR between the two triggers.
I think it would be clear to for us in terms of debugging to see which alarm is triggered since on the low side, temp - 4 would be lower than 33, (31).

Wording may be unclear. Im up for suggestions to help clarify the alarm differences.

Fixed. Thanks!

Fixed. Thanks!

Why alarm if above low temp? I think >42 or > tgt+4 should be two ways to get same non-safety high temp alarm. And I think <33 or < tgt-4 should be two ways to get same non-safety low temp alarm. D...

Why alarm if above low temp? I think >42 or > tgt+4 should be two ways to get same non-safety high temp alarm.
And I think <33 or < tgt-4 should be two ways to get same non-safety low temp alarm.
Did Systems want to separate these? My understanding was not to separate.

Let's review PRS req together. Seems like we have conflicting alarm properties in place that don't make sense to me. Let's figure out what behavior Systems team is looking for here and then decide ...

Let's review PRS req together. Seems like we have conflicting alarm properties in place that don't make sense to me. Let's figure out what behavior Systems team is looking for here and then decide what properties/persistence should be set.

PRS (Ex: PRS 377) no longer has it as of 1/05. SRS (Ex: SRSHD 901) does. which one should be enforced?

PRS (Ex: PRS 377) no longer has it as of 1/05. SRS (Ex: SRSHD 901) does. which one should be enforced?

I thought the non-safety temp alarms had a large persistence. Now they have none?

I thought the non-safety temp alarms had a large persistence. Now they have none?

If you clear immediately, it will let you resume/clear alarm immediately (before temp comes back into range), but then alarm will re-trigger immediately (because temp is still out of range).

If you clear immediately, it will let you resume/clear alarm immediately (before temp comes back into range), but then alarm will re-trigger immediately (because temp is still out of range).

HD-DEN-14830_Incorrect Dialysate Temperature Alarm Triggered
HD-DEN-14830_Incorrect Dialysate Temperature Alarm Triggered
HD-DEN-14830_Incorrect Dialysate Temperature Alarm Triggered
HD-DEN-14830_Incorrect Dialysate Temperature Alarm Triggered