Index: AlarmMapping.csv =================================================================== diff -u -r86e9dfbff50cb7e16fd94c16c1c818cef3b47eac -ra5fe0c9504558ddafc2af2ccd852b8f311a00616 --- AlarmMapping.csv (.../AlarmMapping.csv) (revision 86e9dfbff50cb7e16fd94c16c1c818cef3b47eac) +++ AlarmMapping.csv (.../AlarmMapping.csv) (revision a5fe0c9504558ddafc2af2ccd852b8f311a00616) @@ -104,7 +104,7 @@ 103,"HD non-volatile calibration group invalid record CRC." 104,"Air trap level sensors reporting illegal combination of air/fluid." 105,"DG non-volatile system group invalid record CRC." - 106,"Unused." + 106,"DG dialysate flow sensor invalid calibration record." 107,"HD reports DG restarted fault." 108,"HD syringe pump ADC error." 109,"HD syringe pump volume check error." @@ -114,7 +114,7 @@ 113,"HD blood leak zero command fault." 114,"HD venous air bubble detector self-test failure." 115,"DG temperature sensor out of range." - 116,"DG temperature sensor ADC out of range." + 116,"DG temperature sensor ADC fault." 117,"DG CAN message not acked." 118,"DG RTC config error." 119,"DG RTC battery low error." @@ -142,7 +142,7 @@ 141,"Air trap fill timeout during treatment." 142,"Blood pump occlusion detected." 143,"DG dialysate temperature sensors out of range." - 144,"Alarm ID available for use." + 144,"DG RO flow sensor invalid calibration record." 145,"Concentrate conductivity after adding acid out of range alarm." 146,"DG RTC or timer accuracy failure." 147,"Wait for the DG to produce dialysate." @@ -191,13 +191,13 @@ 190,"DG drain pump off fault." 191,"DG flow rate out of upper range." 192,"DG flow rate out of lower range." - 193,"." + 193,"Alarm available for use." 194,"DG RO pump duty cycle out of range." 195,"DG RO pump pressure out of range." - 196,"DG temperature sensor error flag fault." - 197,"DG temperature sensors ADC fault." + 196,"DG temperature sensor error bit fault." + 197,"DG temperature sensors FPGA fault." 198,"Alarm ID available for use." - 199,"." + 199,"Alarm ID available for use." 200,"DG thermistors/sensors temperature out of range." 201,"HD pre-treatment mode wet self-test lc vs FMD failure." 202,"HD pre-treatment mode dry pressure normal self-test failure."