Index: AlarmMapping.csv =================================================================== diff -u -r86e9dfbff50cb7e16fd94c16c1c818cef3b47eac -rb3a919eb1696079530d537fb2bf8b7e526178997 --- AlarmMapping.csv (.../AlarmMapping.csv) (revision 86e9dfbff50cb7e16fd94c16c1c818cef3b47eac) +++ AlarmMapping.csv (.../AlarmMapping.csv) (revision b3a919eb1696079530d537fb2bf8b7e526178997) @@ -8,25 +8,25 @@ 7,"HD RTC battery low." 8,"HD accelerometer failed POST." 9,"DG accelerometer failed POST." - 10,"Alarm ID available for use." - 11,"Alarm ID available for use." + 10,"HD blood leak sensor setting embedded mode failure." + 11,"HD dialysate temperature below target temperature." 12,"HD firmware image integrity POST test failed." 13,"DG firmware image integrity POST test failed." - 14,"HD blood flow invalid calibration." - 15,"HD dialysate flow invalid calibration." + 14,"DG invalid usage record CRC." + 15,"HD invalid usage record CRC." 16,"HD alarm audio failed POST." 17,"HD UI POST failed." 18,"DG load cells invalid calibration record." - 19,"DG flow sensors invalid calibration record." + 19,"HD dialysate temperature above target temperature." 20,"DG conductivity sensors invalid calibration record." 21,"DG drain line volume invalid calibration record." 22,"DG reservoirs invalid calibration record." 23,"DG acid concentrate calibration record." 24,"DG bicarb concentrate calibration record." 25,"DG accelerometer invalid calibration record." 26,"HD accelerometer invalid calibration record." - 27,"HD blood flow invalid calibration record." - 28,"HD dialyzer flow invalid calibration record." + 27,"HD blood leak sensor zero and self test sequence failed." + 28,"DG two wire sensors FPGA fault." 29,"HD heparin force sensor invalid calibration record." 30,"HD Software fault.\nSoftware found itself in an unexpected state." 31,"Blood pump failed motor controller current check.\nToo high when pump should be off or out of range when pump should be running." @@ -70,8 +70,8 @@ 69,"HD blood leak sensor invalid calibration record." 70,"HD arterial pressure self-test failure alarm." 71,"HD venous pressure self-test failure alarm." - 72,"HD blood flow meter status check self-test failure alarm." - 73,"HD dialysate flow meter status check self-test failure alarm." + 72,"HD No load cell data message receive at least once every 2 seconds." + 73,"HD No dialysate temperature data message receive at least once every 2 seconds." 74,"HD blood leak detector self-test transmit timeout." 75,"HD syringe pump self-test failure alarm." 76,"HD monitored voltage is out of range." @@ -101,10 +101,10 @@ 100,"HD no cartridge self-test timeout." 101,"HD dry self-test timeout." 102,"DG heat disinfect temperature gradient out of range." - 103,"HD non-volatile calibration group invalid record CRC." + 103,"HD invalid calibration CRC." 104,"Air trap level sensors reporting illegal combination of air/fluid." - 105,"DG non-volatile system group invalid record CRC." - 106,"Unused." + 105,"DG invalid calibration CRC." + 106,"DG dialysate flow sensor invalid calibration record." 107,"HD reports DG restarted fault." 108,"HD syringe pump ADC error." 109,"HD syringe pump volume check error." @@ -114,7 +114,7 @@ 113,"HD blood leak zero command fault." 114,"HD venous air bubble detector self-test failure." 115,"DG temperature sensor out of range." - 116,"DG temperature sensor ADC out of range." + 116,"DG temperature sensor ADC fault." 117,"DG CAN message not acked." 118,"DG RTC config error." 119,"DG RTC battery low error." @@ -142,7 +142,7 @@ 141,"Air trap fill timeout during treatment." 142,"Blood pump occlusion detected." 143,"DG dialysate temperature sensors out of range." - 144,"Alarm ID available for use." + 144,"DG RO flow sensor invalid calibration record." 145,"Concentrate conductivity after adding acid out of range alarm." 146,"DG RTC or timer accuracy failure." 147,"Wait for the DG to produce dialysate." @@ -153,8 +153,8 @@ 152,"Inlet water pressure in low range." 153,"HD prime completed high priority alarm." 154,"EEPROM operations (read, write, erase) failure." - 155,"Alarm ID available for use." - 156,"Alarm ID available for use." + 155,"HD No reservoirs data message receive at least once every 2 seconds." + 156,"HD No DG operation mode message receive at least once every 2 seconds." 157,"DG chemical disinfect prime acid line timeout." 158,"Inlet water temperature is in the low range." 159,"Dialysate inlet pump failed flow vs motor speed check." @@ -174,7 +174,7 @@ 173,"HD RTC configuration error." 174,"HD RTC or timer accuracy failure." 175,"HD pump direction status error." - 176,"HD active reservoir is full during treatment and DG not ready to switch reservoirs." + 176,"HD dialysate temperature out of high safety range." 177,"DG software fault.\nSoftware found itself in an unexpected state." 178,"HD communication timeout." 179,"DG FPGA communication down for too long." @@ -191,13 +191,13 @@ 190,"DG drain pump off fault." 191,"DG flow rate out of upper range." 192,"DG flow rate out of lower range." - 193,"." + 193,"HD dialysate temperature out of low safety range." 194,"DG RO pump duty cycle out of range." 195,"DG RO pump pressure out of range." - 196,"DG temperature sensor error flag fault." - 197,"DG temperature sensors ADC fault." - 198,"Alarm ID available for use." - 199,"." + 196,"DG temperature sensor error bit fault." + 197,"DG temperature sensors FPGA fault." + 198,"RO flow too low while primary heater is on." + 199,"Dialysate flow too low while trimmer heater is on." 200,"DG thermistors/sensors temperature out of range." 201,"HD pre-treatment mode wet self-test lc vs FMD failure." 202,"HD pre-treatment mode dry pressure normal self-test failure." @@ -206,7 +206,7 @@ 205,"Conductivity sensor fault." 206,"DG dialysate fill runs out of time." 207,"DG flow meter check failure alarm." - 208,"Alarm ID available for use." + 208,"HD venous pressure occlusion alarm." 209,"DG drain circulation line timeout." 210,"HD battery pack detected an error." 211,"Blood sitting too long warning (>4 min)." @@ -218,20 +218,20 @@ 217,"HD blood prime volume check failure." 218,"HD syringe detected alarm." 219,"HD syringe pump syringe removed alarm." - 220,"HD battery pack SOC is too low." + 220,"DG THd sensors FPGA fault." 221,"HD Saline bag is empty." 222,"HD dialysate temperature too high alarm." 223,"HD dialysate temperature too low alarm." 224,"Treatment stopped by user action." 225,"HD end of treatment warning." 226,"HD prime completed low priority alarm." 227,"HD prime out of time alarm." - 228,"HD time out on prime sub-mode purge air state." + 228,"HD time out on prime saline purge air state." 229,"HD prime dialysate dialyzer time out alarm." 230,"HD prime dialysate bypass time out alarm." 231,"HD pre-treatment mode dry pressure self-test failure." 232,"HD pre-treatment mode wet self-test LC vs LC failure." - 233,"Alarm ID available for use." + 233,"DG barometric sensor coefficients bad CRC." 234,"RTC (i.e read time) or RAM operations failure (read or write)." 235,"HD in treatment stopped sub-mode after rinseback completed (no escalation)." 236,"HD needs new cartridge to be installed." @@ -289,4 +289,16 @@ 288,"HD load cells primary/back up drift out of range." 289,"DG dialysate or concentrate caps not closed." 290,"DG flow too low while heater is on." - 291,"Total number of alarms." + 291,"HD processor is in Disinfect Flush mode." + 292,"HD processor is in Disinfect Heat mode." + 293,"HD processor is in Disinfect Chemical mode." + 294,"HD processor Disinfect Chemical mode instructions to user." + 295,"UI POST CloudSync failure." + 296,"DG TDi sensors FPGA fault." + 297,"DG TRo sensors FPGA fault." + 298,"DG baro sensor FPGA fault." + 299,"DG disinfect has been expired." + 300,"DG service time interval has elapsed." + 301,"DG primary heater power voltage out of range." + 302,"DG processor RAM error." + 303,"Total number of alarms."