Index: AlarmMapping.csv =================================================================== diff -u -r99e5b87f73c3b849ba33a212ec84cf4bf6ef38f1 -r1da89b0452b8ef9448847618e75c118f3f58bd0c --- AlarmMapping.csv (.../AlarmMapping.csv) (revision 99e5b87f73c3b849ba33a212ec84cf4bf6ef38f1) +++ AlarmMapping.csv (.../AlarmMapping.csv) (revision 1da89b0452b8ef9448847618e75c118f3f58bd0c) @@ -283,7 +283,7 @@ 282,"Bicarbonate concentration bottle low volume alarm." 283,"DG load cells weight out of range." 284,"DG load cells primary/back up drift out of range." - 285,"Available for use." + 285,"HD processor is in RO permeate sample mode." 286,"DG concentrate cap not in proper position." 287,"HD processor clock speed checks against FPGA clock failure." 288,"HD load cells primary/back up drift out of range." @@ -292,14 +292,14 @@ 291,"HD processor is in Disinfect Flush mode." 292,"HD processor is in Disinfect Heat mode." 293,"HD processor is in Disinfect Chemical mode." - 294,"Available for use." + 294,"HD processor is in Disinfect active cool mode." 295,"HD UI POST CloudSync failure." 296,"DG TDi sensors FPGA fault." 297,"DG TRo sensors FPGA fault." 298,"DG baro sensor FPGA fault." 299,"DG invalid serial number." 300,"HD invalid serial number." - 301,"Alarm ID available." + 301,"DG RO permeate sample remove the dialysate cap." 302,"DG processor RAM error." 303,"DG drain pump direction FPGA fault." 304,"HD invalid usage record."