Index: AlarmMapping.csv =================================================================== diff -u -r9ae3b0d6624904693329309aaf8ff02784c17184 -r3779eaac00ecfa03d5b0405171ab68a0e0ff5538 --- AlarmMapping.csv (.../AlarmMapping.csv) (revision 9ae3b0d6624904693329309aaf8ff02784c17184) +++ AlarmMapping.csv (.../AlarmMapping.csv) (revision 3779eaac00ecfa03d5b0405171ab68a0e0ff5538) @@ -85,17 +85,17 @@ 84,"HD venous pressure sensor not being read." 85,"HD venous pressure sensor temperature out of range error." 86,"HD BP occlusion sensor not being read." - 87,"HD DPi occlusion sensor not being read." - 88,"HD DPo occlusion sensor not being read." + 87,"This alarm ID is available for use." + 88,"This alarm ID is available for use." 89,"HD BP occlusion sensor error." - 90,"HD DPi occlusion sensor error." - 91,"HD DPo occlusion sensor error." - 92,"HD blood flow sensor not being read." - 93,"HD blood flow sensor error." - 94,"HD dialysate flow sensor not being read." - 95,"HD dialysate flow sensor error." - 96,"HD blood flow sensor not being read - slow packet." - 97,"HD dialysate flow sensor not being read - slow packet." + 90,"This alarm ID is available for use." + 91,"HD No dialysate flow data receive in the last 3 seconds." + 92,"This alarm ID is available for use." + 93,"This alarm ID is available for use." + 94,"This alarm ID is available for use." + 95,"This alarm ID is available for use." + 96,"This alarm ID is available for use." + 97,"This alarm ID is available for use." 98,"HD battery communication fault." 99,"HD syringe pump stall alarm." 100,"HD no cartridge self-test timeout." @@ -277,7 +277,7 @@ 276,"HD venous pressure sensor is reading out of range." 277,"HD BP occlusion sensor is reading out of range." 278,"This alarm ID is available for use." - 279,"This alarm ID is available for use." + 279,"DG dialysate drain time out." 280,"HD arterial pressure sensor read timeout error." 281,"Acid concentration bottle low volume alarm." 282,"Bicarbonate concentration bottle low volume alarm."