Index: AlarmMapping.csv =================================================================== diff -u -ra5fe0c9504558ddafc2af2ccd852b8f311a00616 -r51bfd04e3f843dd29f7171a678eb000434efd18c --- AlarmMapping.csv (.../AlarmMapping.csv) (revision a5fe0c9504558ddafc2af2ccd852b8f311a00616) +++ AlarmMapping.csv (.../AlarmMapping.csv) (revision 51bfd04e3f843dd29f7171a678eb000434efd18c) @@ -8,25 +8,25 @@ 7,"HD RTC battery low." 8,"HD accelerometer failed POST." 9,"DG accelerometer failed POST." - 10,"Alarm ID available for use." - 11,"Alarm ID available for use." + 10,"Not used." + 11,"HD dialysate temperature below target temperature." 12,"HD firmware image integrity POST test failed." 13,"DG firmware image integrity POST test failed." - 14,"HD blood flow invalid calibration." - 15,"HD dialysate flow invalid calibration." + 14,"DG invalid usage record CRC." + 15,"HD invalid usage record CRC." 16,"HD alarm audio failed POST." 17,"HD UI POST failed." 18,"DG load cells invalid calibration record." - 19,"DG flow sensors invalid calibration record." + 19,"HD dialysate temperature above target temperature." 20,"DG conductivity sensors invalid calibration record." 21,"DG drain line volume invalid calibration record." 22,"DG reservoirs invalid calibration record." 23,"DG acid concentrate calibration record." 24,"DG bicarb concentrate calibration record." 25,"DG accelerometer invalid calibration record." 26,"HD accelerometer invalid calibration record." - 27,"HD blood flow invalid calibration record." - 28,"HD dialyzer flow invalid calibration record." + 27,"Not used." + 28,"Not used." 29,"HD heparin force sensor invalid calibration record." 30,"HD Software fault.\nSoftware found itself in an unexpected state." 31,"Blood pump failed motor controller current check.\nToo high when pump should be off or out of range when pump should be running." @@ -70,8 +70,8 @@ 69,"HD blood leak sensor invalid calibration record." 70,"HD arterial pressure self-test failure alarm." 71,"HD venous pressure self-test failure alarm." - 72,"HD blood flow meter status check self-test failure alarm." - 73,"HD dialysate flow meter status check self-test failure alarm." + 72,"Not used." + 73,"Not used." 74,"HD blood leak detector self-test transmit timeout." 75,"HD syringe pump self-test failure alarm." 76,"HD monitored voltage is out of range." @@ -101,9 +101,9 @@ 100,"HD no cartridge self-test timeout." 101,"HD dry self-test timeout." 102,"DG heat disinfect temperature gradient out of range." - 103,"HD non-volatile calibration group invalid record CRC." + 103,"HD invalid calibration CRC." 104,"Air trap level sensors reporting illegal combination of air/fluid." - 105,"DG non-volatile system group invalid record CRC." + 105,"DG invalid calibration CRC." 106,"DG dialysate flow sensor invalid calibration record." 107,"HD reports DG restarted fault." 108,"HD syringe pump ADC error." @@ -153,8 +153,8 @@ 152,"Inlet water pressure in low range." 153,"HD prime completed high priority alarm." 154,"EEPROM operations (read, write, erase) failure." - 155,"Alarm ID available for use." - 156,"Alarm ID available for use." + 155,"Not used." + 156,"Not used." 157,"DG chemical disinfect prime acid line timeout." 158,"Inlet water temperature is in the low range." 159,"Dialysate inlet pump failed flow vs motor speed check." @@ -174,7 +174,7 @@ 173,"HD RTC configuration error." 174,"HD RTC or timer accuracy failure." 175,"HD pump direction status error." - 176,"HD active reservoir is full during treatment and DG not ready to switch reservoirs." + 176,"HD dialysate temperature out of high safety range." 177,"DG software fault.\nSoftware found itself in an unexpected state." 178,"HD communication timeout." 179,"DG FPGA communication down for too long." @@ -191,13 +191,13 @@ 190,"DG drain pump off fault." 191,"DG flow rate out of upper range." 192,"DG flow rate out of lower range." - 193,"Alarm available for use." + 193,"HD dialysate temperature out of low safety range." 194,"DG RO pump duty cycle out of range." 195,"DG RO pump pressure out of range." 196,"DG temperature sensor error bit fault." 197,"DG temperature sensors FPGA fault." - 198,"Alarm ID available for use." - 199,"Alarm ID available for use." + 198,"RO flow too low while primary heater is on." + 199,"Dialysate flow too low while trimmer heater is on." 200,"DG thermistors/sensors temperature out of range." 201,"HD pre-treatment mode wet self-test lc vs FMD failure." 202,"HD pre-treatment mode dry pressure normal self-test failure." @@ -206,7 +206,7 @@ 205,"Conductivity sensor fault." 206,"DG dialysate fill runs out of time." 207,"DG flow meter check failure alarm." - 208,"Alarm ID available for use." + 208,"HD venous pressure occlusion alarm." 209,"DG drain circulation line timeout." 210,"HD battery pack detected an error." 211,"Blood sitting too long warning (>4 min)." @@ -231,7 +231,7 @@ 230,"HD prime dialysate bypass time out alarm." 231,"HD pre-treatment mode dry pressure self-test failure." 232,"HD pre-treatment mode wet self-test LC vs LC failure." - 233,"Alarm ID available for use." + 233,"DG barometric sensor coefficients bad CRC." 234,"RTC (i.e read time) or RAM operations failure (read or write)." 235,"HD in treatment stopped sub-mode after rinseback completed (no escalation)." 236,"HD needs new cartridge to be installed." @@ -289,4 +289,8 @@ 288,"HD load cells primary/back up drift out of range." 289,"DG dialysate or concentrate caps not closed." 290,"DG flow too low while heater is on." - 291,"Total number of alarms." + 291,"HD processor is in Disinfect Flush mode." + 292,"HD processor is in Disinfect Heat mode." + 293,"HD processor is in Disinfect Chemical mode." + 294,"HD processor Disinfect Chemical mode instructions to user." + 295,"Total number of alarms."