Index: AlarmMapping.csv =================================================================== diff -u -rc4b95505c1d02a74d9c1b8132578bbd8a21cec9a -rb238ed3e1aa94f5c03d4e1a6ed6c9fdfdd2a3cb5 --- AlarmMapping.csv (.../AlarmMapping.csv) (revision c4b95505c1d02a74d9c1b8132578bbd8a21cec9a) +++ AlarmMapping.csv (.../AlarmMapping.csv) (revision b238ed3e1aa94f5c03d4e1a6ed6c9fdfdd2a3cb5) @@ -8,7 +8,7 @@ 7,"HD RTC battery low." 8,"HD accelerometer failed POST." 9,"DG accelerometer failed POST." - 10,"Not used." + 10,"HD blood leak sensor setting embedded mode failure." 11,"HD dialysate temperature below target temperature." 12,"HD firmware image integrity POST test failed." 13,"DG firmware image integrity POST test failed." @@ -25,8 +25,8 @@ 24,"DG bicarb concentrate calibration record." 25,"DG accelerometer invalid calibration record." 26,"HD accelerometer invalid calibration record." - 27,"Not used." - 28,"Not used." + 27,"HD blood leak sensor zero and self test sequence failed." + 28,"DG two wire sensors FPGA fault." 29,"HD heparin force sensor invalid calibration record." 30,"HD Software fault.\nSoftware found itself in an unexpected state." 31,"Blood pump failed motor controller current check.\nToo high when pump should be off or out of range when pump should be running." @@ -70,8 +70,8 @@ 69,"HD blood leak sensor invalid calibration record." 70,"HD arterial pressure self-test failure alarm." 71,"HD venous pressure self-test failure alarm." - 72,"Not used." - 73,"Not used." + 72,"HD No load cell data message receive at least once every 2 seconds." + 73,"HD No dialysate temperature data message receive at least once every 2 seconds." 74,"HD blood leak detector self-test transmit timeout." 75,"HD syringe pump self-test failure alarm." 76,"HD monitored voltage is out of range." @@ -153,8 +153,8 @@ 152,"Inlet water pressure in low range." 153,"HD prime completed high priority alarm." 154,"EEPROM operations (read, write, erase) failure." - 155,"Not used." - 156,"Not used." + 155,"HD No reservoirs data message receive at least once every 2 seconds." + 156,"HD No DG operation mode message receive at least once every 2 seconds." 157,"DG chemical disinfect prime acid line timeout." 158,"Inlet water temperature is in the low range." 159,"Dialysate inlet pump failed flow vs motor speed check." @@ -218,15 +218,15 @@ 217,"HD blood prime volume check failure." 218,"HD syringe detected alarm." 219,"HD syringe pump syringe removed alarm." - 220,"HD battery pack SOC is too low." + 220,"DG THd sensors FPGA fault." 221,"HD Saline bag is empty." 222,"HD dialysate temperature too high alarm." 223,"HD dialysate temperature too low alarm." 224,"Treatment stopped by user action." 225,"HD end of treatment warning." 226,"HD prime completed low priority alarm." 227,"HD prime out of time alarm." - 228,"HD time out on prime sub-mode purge air state." + 228,"HD time out on prime saline purge air state." 229,"HD prime dialysate dialyzer time out alarm." 230,"HD prime dialysate bypass time out alarm." 231,"HD pre-treatment mode dry pressure self-test failure." @@ -293,4 +293,8 @@ 292,"HD processor is in Disinfect Heat mode." 293,"HD processor is in Disinfect Chemical mode." 294,"HD processor Disinfect Chemical mode instructions to user." - 295,"Total number of alarms." + 295,"UI POST CloudSync failure." + 296,"DG TDi sensors FPGA fault." + 297,"DG TRo sensors FPGA fault." + 298,"DG baro sensor FPGA fault." + 299,"Total number of alarms."