Index: AlarmMapping.csv =================================================================== diff -u -rdf0329c47627ceb1fe460cda2ae3ff8c9c851106 -r0f1aa395dfb46e006bdec10d0bb99f2ed27739bf --- AlarmMapping.csv (.../AlarmMapping.csv) (revision df0329c47627ceb1fe460cda2ae3ff8c9c851106) +++ AlarmMapping.csv (.../AlarmMapping.csv) (revision 0f1aa395dfb46e006bdec10d0bb99f2ed27739bf) @@ -91,7 +91,7 @@ 90,"This alarm ID is available for use." 91,"HD No dialysate flow data receive in the last 3 seconds." 92,"DG conductivity sensor bad status." - 93,"This alarm ID is available for use." + 93,"DG outlet primary conductivity out of range." 94,"This alarm ID is available for use." 95,"This alarm ID is available for use." 96,"Inlet water temperature in high range." @@ -141,7 +141,7 @@ 140,"Dialysate generator communication timeout." 141,"Air trap fill timeout during treatment." 142,"Blood pump occlusion detected." - 143,"DG dialysate temperature sensors drift timeout." + 143,"DG dialysate temperature sensors out of range." 144,"DG software configuration record invalid CRC." 145,"Concentrate conductivity after adding acid out of range alarm." 146,"Dialysate conductivity out of range." @@ -199,7 +199,7 @@ 198,"DG heaters cold junction temperature out of range." 199,"DG heaters fault." 200,"DG thermistors/sensors temperature out of range." - 201,"Inlet water quality is bad (temperature, conductivity, pressure)." + 201,"This alarm ID is available for use." 202,"Inlet water pressure fault." 203,"Pressure sensor fault." 204,"RO rejection ratio out of range."