Index: AlarmMapping.csv =================================================================== diff -u -refb549a694539630ba7c1d8e348f598052b85f0f -rb238ed3e1aa94f5c03d4e1a6ed6c9fdfdd2a3cb5 --- AlarmMapping.csv (.../AlarmMapping.csv) (revision efb549a694539630ba7c1d8e348f598052b85f0f) +++ AlarmMapping.csv (.../AlarmMapping.csv) (revision b238ed3e1aa94f5c03d4e1a6ed6c9fdfdd2a3cb5) @@ -26,7 +26,7 @@ 25,"DG accelerometer invalid calibration record." 26,"HD accelerometer invalid calibration record." 27,"HD blood leak sensor zero and self test sequence failed." - 28,"Not used." + 28,"DG two wire sensors FPGA fault." 29,"HD heparin force sensor invalid calibration record." 30,"HD Software fault.\nSoftware found itself in an unexpected state." 31,"Blood pump failed motor controller current check.\nToo high when pump should be off or out of range when pump should be running." @@ -218,7 +218,7 @@ 217,"HD blood prime volume check failure." 218,"HD syringe detected alarm." 219,"HD syringe pump syringe removed alarm." - 220,"Not used." + 220,"DG THd sensors FPGA fault." 221,"HD Saline bag is empty." 222,"HD dialysate temperature too high alarm." 223,"HD dialysate temperature too low alarm." @@ -294,4 +294,7 @@ 293,"HD processor is in Disinfect Chemical mode." 294,"HD processor Disinfect Chemical mode instructions to user." 295,"UI POST CloudSync failure." - 296,"Total number of alarms." + 296,"DG TDi sensors FPGA fault." + 297,"DG TRo sensors FPGA fault." + 298,"DG baro sensor FPGA fault." + 299,"Total number of alarms."