Index: AlarmDefs.h =================================================================== diff -u -r17e260f451a5bc53c4d3d0724cfed4d32a1fb92c -r32a9bd86f5789674eaa4eb150cd56919485e090e --- AlarmDefs.h (.../AlarmDefs.h) (revision 17e260f451a5bc53c4d3d0724cfed4d32a1fb92c) +++ AlarmDefs.h (.../AlarmDefs.h) (revision 32a9bd86f5789674eaa4eb150cd56919485e090e) @@ -638,7 +638,7 @@ { ALARM_PRIORITY_HIGH, 799, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_CONCENTRATE_PUMP_FAULT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_CP1_SPEED_CONTROL_ERROR, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_CP2_SPEED_CONTROL_ERROR, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DRAIN_PUMP_RPM_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DRAIN_PUMP_RPM_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DRAIN_PUMP_OFF_FAULT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_FLOW_RATE_OUT_OF_UPPER_RANGE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_FLOW_RATE_OUT_OF_LOWER_RANGE, 0, ALARM_ID_NO_ALARM }, @@ -907,6 +907,7 @@ { 3, ALARM_ID_DG_RESERVOIR_FILL_TIMEOUT }, { 3, ALARM_ID_DG_INVALID_LOAD_CELL_VALUE }, { 3, ALARM_ID_DG_CHEM_DISINFECT_PRIME_ACID_LINE_TIME_OUT }, + { 3, ALARM_ID_DRAIN_PUMP_RPM_OUT_OF_RANGE }, { 3, ALARM_ID_DG_RESERVOIR_LEAK_TIMEOUT }, { 3, ALARM_ID_DG_CHEM_DISINFECT_TARGET_TEMP_TIMEOUT }, { 4, ALARM_ID_END_OF_TREATMENT_HIGH }, @@ -972,7 +973,6 @@ { 799, ALARM_ID_DG_CONCENTRATE_PUMP_FAULT }, { 799, ALARM_ID_CP1_SPEED_CONTROL_ERROR }, { 799, ALARM_ID_CP2_SPEED_CONTROL_ERROR }, - { 799, ALARM_ID_DRAIN_PUMP_RPM_OUT_OF_RANGE }, { 799, ALARM_ID_DRAIN_PUMP_OFF_FAULT }, { 799, ALARM_ID_FLOW_RATE_OUT_OF_UPPER_RANGE }, { 799, ALARM_ID_FLOW_RATE_OUT_OF_LOWER_RANGE },