Index: AlarmDefs.h =================================================================== diff -u -r2c54b5f20c532089bd6854e15d835efb7cb021c1 -r4effb800b3fd04d9270724d2f439adc42b5f6548 --- AlarmDefs.h (.../AlarmDefs.h) (revision 2c54b5f20c532089bd6854e15d835efb7cb021c1) +++ AlarmDefs.h (.../AlarmDefs.h) (revision 4effb800b3fd04d9270724d2f439adc42b5f6548) @@ -249,13 +249,13 @@ ALARM_ID_DG_CHEM_DISINFECT_TARGET_TEMP_TIMEOUT = 211, ///< DG chemical disinfect could not reach to target temperature ALARM_ID_DG_FPGA_POST_TEST_FAILED = 212, ///< DG FPGA POST failure ALARM_ID_HD_ARTERIAL_BUBBLE_SELF_TEST_FAILURE = 213, ///< HD arterial air bubble detector self-test failure - ALARM_ID_HD_ARTERIAL_BUBBLE_DETECTOR_FAULT = 214, ///< HD arterial air bubble detector fault + ALARM_ID_HD_ARTERIAL_BUBBLE_DETECTED_RINSEBACK = 214, ///< HD arterial air bubble detected rinseback alarm ALARM_ID_DG_HEATERS_SELF_TEST_FAILURE = 215, ///< DG heaters failed POST ALARM_ID_HD_VENOUS_BUBBLE_SELF_TEST_FAILURE = 216, ///< HD venous air bubble detector self-test failure - ALARM_ID_HD_VENOUS_BUBBLE_DETECTOR_TREATMENT_FAULT = 217, ///< HD venous air bubble detector treatment fault - ALARM_ID_HD_VENOUS_BUBBLE_DETECTOR_RINSEBACK_FAULT = 218, ///< HD venous air bubble detector rinseback fault - ALARM_ID_HD_ARTERIAL_BUBBLE_DETECTED = 219, ///< HD arterial bubble detected alarm - ALARM_ID_HD_VENOUS_BUBBLE_DETECTED = 220, ///< HD venous bubble detected alarm + ALARM_ID_AVAILABLE_6 = 217, ///< HD venous air bubble detector treatment fault + ALARM_ID_HD_VENOUS_BUBBLE_DETECTED_RINSEBACK = 218, ///< HD venous air bubble detected rinseback alarm + ALARM_ID_HD_ARTERIAL_BUBBLE_DETECTED = 219, ///< HD arterial air bubble detected alarm + ALARM_ID_HD_VENOUS_BUBBLE_DETECTED = 220, ///< HD venous air bubble detected alarm ALARM_ID_HD_BP_FLOW_SLOW_READ_TIMEOUT_ERROR = 221, ///< HD blood flow sensor not being read - slow packet ALARM_ID_HD_DP_FLOW_SLOW_READ_TIMEOUT_ERROR = 222, ///< HD dialysate flow sensor not being read - slow packet ALARM_ID_HD_PUMP_DIRECTION_STATUS_ERROR = 223, ///< HD pump direction status error @@ -591,11 +591,11 @@ { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_DG_CHEM_DISINFECT_TARGET_TEMP_TIMEOUT { ALARM_PRIORITY_HIGH, 1, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_DG_FPGA_POST_TEST_FAILED { ALARM_PRIORITY_HIGH, 9, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_HD_ARTERIAL_BUBBLE_SELF_TEST_FAILURE - { ALARM_PRIORITY_HIGH, 9, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE, TRUE, TRUE , TRUE , TRUE , TRUE , TRUE, FALSE }, // ALARM_ID_HD_ARTERIAL_BUBBLE_DETECTOR_FAULT + { ALARM_PRIORITY_HIGH, 9, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE, TRUE, TRUE , TRUE , TRUE , TRUE , TRUE, FALSE }, // ALARM_ID_HD_ARTERIAL_BUBBLE_DETECTED_RINSEBACK { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_DG_HEATERS_SELF_TEST_FAILURE { ALARM_PRIORITY_HIGH, 9, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_HD_VENOUS_BUBBLE_SELF_TEST_FAILURE - { ALARM_PRIORITY_HIGH, 9, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE, TRUE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_HD_VENOUS_BUBBLE_DETECTOR_TREATMENT_FAULT - { ALARM_PRIORITY_HIGH, 10, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE, TRUE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_HD_VENOUS_BUBBLE_DETECTOR_RINSEBACK_FAULT + { ALARM_PRIORITY_HIGH, 9, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE, TRUE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_AVAILABLE_6 + { ALARM_PRIORITY_HIGH, 10, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE, TRUE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_HD_VENOUS_BUBBLE_DETECTED_RINSEBACK { ALARM_PRIORITY_HIGH, 9, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE, TRUE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE }, // ALARM_ID_HD_ARTERIAL_BUBBLE_DETECTED { ALARM_PRIORITY_HIGH, 9, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE, TRUE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE }, // ALARM_ID_HD_VENOUS_BUBBLE_DETECTED { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_HD_BP_FLOW_SLOW_READ_TIMEOUT_ERROR