Index: AlarmDefs.h =================================================================== diff -u -r3866a71882e70f5afb1e6e94865522f35c05388f -r7a57d9836f752e34f98e6a359e85b82cb15a0dd0 --- AlarmDefs.h (.../AlarmDefs.h) (revision 3866a71882e70f5afb1e6e94865522f35c05388f) +++ AlarmDefs.h (.../AlarmDefs.h) (revision 7a57d9836f752e34f98e6a359e85b82cb15a0dd0) @@ -8,7 +8,7 @@ * @file AlarmDefs.h * * @author (last) Dara Navaei -* @date (last) 15-Feb-2022 +* @date (last) 30-Mar-2022 * * @author (original) Sean * @date (original) 11-Feb-2020 @@ -125,11 +125,11 @@ ALARM_ID_DG_HEATING_INVALID_CAL_RECORD = 87, ///< DG heating invalid calibration record ALARM_ID_DG_CONC_PUMP_HALL_SENSOR_OUT_OF_RANGE = 88, ///< DG concentrate pumps hall sensor out of range ALARM_ID_HD_BP_OCCLUSION_SENSOR_ERROR = 89, ///< HD BP occlusion sensor error - ALARM_ID____AVAILABLE_25 = 90, ///< This alarm ID is available for use + ALARM_ID_DG_LOAD_CELL_FPGA_READ_ERROR = 90, ///< DG load cells FPGA read error. ALARM_ID_HD_DIALYSATE_FLOW_DATA_NOT_RECEIVE = 91, ///< HD No dialysate flow data receive in the last 3 seconds ALARM_ID_DG_CONDUCTIVITY_SENSOR_BAD_STATUS = 92, ///< DG conductivity sensor bad status. - ALARM_ID____AVAILABLE_15 = 93, ///< This alarm ID is available for use - ALARM_ID____AVAILABLE_16 = 94, ///< This alarm ID is available for use + ALARM_ID_DG_OUTLET_PRIMARY_CONDUCTIVITY_OUT_OF_RANGE = 93, ///< DG outlet primary conductivity out of range + ALARM_ID_DG_PRESSURE_OUT_OF_RANGE = 94, ///< DG pressure out of range ALARM_ID____AVAILABLE_17 = 95, ///< This alarm ID is available for use ALARM_ID_INLET_WATER_TEMPERATURE_IN_HIGH_RANGE = 96, ///< Inlet water temperature in high range ALARM_ID_FILL_CONDUCTIVITY_OUT_OF_RANGE = 97, ///< Fill conductivity out of range @@ -152,10 +152,10 @@ ALARM_ID_HD_VENOUS_BUBBLE_SELF_TEST_FAILURE = 114, ///< HD venous air bubble detector self-test failure ALARM_ID_DG_TEMPERATURE_SENSOR_OUT_OF_RANGE = 115, ///< DG temperature sensor out of range ALARM_ID_DG_TEMPERATURE_SENSOR_ADC_OUT_OF_RANGE = 116, ///< DG temperature sensor ADC out of range - ALARM_ID_DG_PRIMARY_HEATER_INTERNAL_TEMP_OUT_OF_RANGE = 117, ///< DG primary heater internal temperature out of range // TODO remove when DEN-11750 was merged - ALARM_ID_DG_PRIMARY_HEATER_CJ_TEMP_OUT_OF_RANGE = 118, ///< DG primary heater cold junction temperature out of range // TODO remove when DEN-11750 was merged - ALARM_ID_DG_TRIMMER_HEATER_INTERNAL_TEMP_OUT_OF_RANGE = 119, ///< DG trimmer heater internal temperature out of range // TODO remove when DEN-11750 was merged - ALARM_ID_DG_TRIMMER_HEATER_CJ_TEMP_OUT_OF_RANGE = 120, ///< DG trimmer heater cold junction temperature out of range // TODO remove when DEN-11750 was merged + ALARM_ID____AVAILABLE_33 = 117, ///< This alarm ID is available for use + ALARM_ID____AVAILABLE_32 = 118, ///< This alarm ID is available for use + ALARM_ID____AVAILABLE_31 = 119, ///< This alarm ID is available for use + ALARM_ID____AVAILABLE_30 = 120, ///< This alarm ID is available for use ALARM_ID_DG_MAIN_PRIMARY_HEATER_VOLTAGE_OUT_OF_RANGE = 121, ///< DG main primary heater voltage out of range ALARM_ID_DG_SMALL_PRIMARY_HEATER_VOLTAGE_OUT_OF_RANGE = 122, ///< DG small primary heater voltage out of range ALARM_ID_DG_TRIMMER_HEATER_VOLTAGE_OUT_OF_RANGE = 123, ///< DG trimmer heater voltage out of range @@ -178,10 +178,10 @@ ALARM_ID_DG_COMM_TIMEOUT = 140, ///< Dialysate generator communication timeout ALARM_ID_AIR_TRAP_FILL_DURING_TREATMENT = 141, ///< Air trap fill timeout during treatment ALARM_ID_OCCLUSION_BLOOD_PUMP = 142, ///< Blood pump occlusion detected - ALARM_ID_DG_DIALYSATE_TEMPERATURE_SENSORS_DRIFT_TIMEOUT = 143, ///< DG dialysate temperature sensors drift timeout. + ALARM_ID_DG_DIALYSATE_TEMPERATURE_SENSORS_OUT_OF_RANGE = 143, ///< DG dialysate temperature sensors out of range ALARM_ID_DG_SW_CONFIG_RECORD_INVALID_CRC = 144, ///< DG software configuration record invalid CRC. ALARM_ID_ACID_CONDUCTIVITY_OUT_OF_RANGE = 145, ///< Concentrate conductivity after adding acid out of range alarm - ALARM_ID_DIALYSATE_CONDUCTIVITY_OUT_OF_RANGE = 146, ///< Dialysate conductivity out of range + ALARM_ID____AVAILABLE_34 = 146, ///< This alarm ID is available for use ALARM_ID_CREATING_DIALYSATE_PLEASE_WAIT = 147, ///< Wait for the DG to produce dialysate ALARM_ID_INLET_WATER_HIGH_TEMPERATURE = 148, ///< Inlet water temperature too high ALARM_ID____AVAILABLE_28 = 149, ///< Alarm ID available for use @@ -233,11 +233,11 @@ ALARM_ID_RO_PUMP_PRESSURE_OUT_OF_RANGE = 195, ///< DG RO pump pressure out of range ALARM_ID_DG_TEMPERATURE_SENSOR_FAULT = 196, ///< DG temperature sensor error flag fault. ALARM_ID_DG_TEMPERATURE_SENSORS_ADC_FAULT = 197, ///< DG temperature sensors ADC fault - ALARM_ID_DG_HEATERS_NEGATIVE_COLD_JUNCTION_TEMPERATURE = 198, ///< DG heaters cold junction temperature out of range + ALARM_ID_DG_HEATERS_NEGATIVE_COLD_JUNCTION_TEMPERATURE = 198, ///< DG heaters cold junction temperature out of range // TODO remove when DEN-12224 merged ALARM_ID_DG_HEATERS_FAULT = 199, ///< DG heaters fault ALARM_ID_DG_THERMISTORS_TEMPERATURE_OUT_OF_RANGE = 200, ///< DG thermistors/sensors temperature out of range - ALARM_ID_DG_BAD_INLET_WATER_QUALITY = 201, ///< Inlet water quality is bad (temperature, conductivity, pressure) - ALARM_ID_INLET_WATER_PRESSURE_FAULT = 202, ///< Inlet water pressure fault + ALARM_ID____AVAILABLE_29 = 201, ///< This alarm ID is available for use + ALARM_ID____AVAILABLE_35 = 202, ///< This alarm ID is available for use ALARM_ID_PRESSURE_SENSOR_FAULT = 203, ///< Pressure sensor fault ALARM_ID_RO_REJECTION_RATIO_OUT_OF_RANGE = 204, ///< RO rejection ratio out of range ALARM_ID_CONDUCTIVITY_SENSOR_FAULT = 205, ///< Conductivity sensor fault @@ -398,7 +398,7 @@ ALARM_STATE_FLAG_BIT_POS_LAMP_ON = 10, ///< Alarm lamp is currently on (for syncing to UI) ALARM_STATE_FLAG_BIT_POS_UNUSED_1 = 11, ///< Reserved ALARM_STATE_FLAG_BIT_POS_UNUSED_2 = 12, ///< Reserved - ALARM_STATE_FLAG_BIT_POS_UNUSED_3 = 13, ///< Reserved + ALARM_STATE_FLAG_BIT_POS_DIALYSATE_RECIRC = 13, ///< Dialysate recirculation while treatment stopped is disabled ALARM_STATE_FLAG_BIT_POS_NO_MINIMIZE = 14, ///< Prevent user from minimizing alarm window ALARM_STATE_FLAG_BIT_POS_TOP_CONDITION = 15, ///< The top alarm's condition is still being detected NUM_OF_ALARM_STATE_FLAG_BIT_POS = 16 ///< The number of alarm state flag bit positions @@ -450,8 +450,8 @@ { ALARM_PRIORITY_HIGH, 1, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_WATCHDOG_POST_TEST_FAILED, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_UI_COMM_POST_FAILED, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 1, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_RTC_CONFIG_ERROR, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_ACCELEROMETER_SELF_TEST_FAILURE, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 1, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_ACCELEROMETER_SELF_TEST_FAILURE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_ACCELEROMETER_SELF_TEST_FAILURE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 1, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_ACCELEROMETER_SELF_TEST_FAILURE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 1, TRUE , TRUE , TRUE , TRUE , TRUE, TRUE, TRUE, TRUE, TRUE , FALSE, FALSE, ALARM_ID_RTC_OR_TIMER_ACCURACY_FAILURE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 1, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_HEATERS_SELF_TEST_FAILURE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, FALSE, FALSE, ALARM_ID_HD_INTEGRITY_POST_TEST_FAILED, 0, ALARM_ID_NO_ALARM }, @@ -522,37 +522,37 @@ // Priority Rank Fault DGFault Stops ClrIm NoClr NoRes NoRin NoEnd NoRcr ClrOnly TxLog ID Escalate In Escalate To { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_SYRINGE_PUMP_SELF_TEST_FAILURE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_VOLTAGE_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_VOLTAGE_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_VOLTAGE_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_SYRINGE_PUMP_ENCODER_DIRECTION_ERROR, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_SYRINGE_PUMP_CONTROLLER_DIRECTION_ERROR, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_SYRINGE_PUMP_FAULT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_SYRINGE_PUMP_OVER_TRAVEL_ERROR, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_SYRINGE_PUMP_DAC_WRITE_ERROR, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_SYRINGE_PUMP_RUNNING_WHILE_BP_OFF_ERROR, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_VENOUS_PRESSURE_READ_TIMEOUT_ERROR, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_PUMP_TRACK_LATCH_OPENED, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_PUMP_TRACK_LATCH_OPENED, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_BP_OCCLUSION_READ_TIMEOUT_ERROR, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_HEATING_INVALID_CAL_RECORD, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_CONC_PUMP_HALL_SENSOR_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_BP_OCCLUSION_SENSOR_ERROR, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID____AVAILABLE_25, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_CONC_PUMP_HALL_SENSOR_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_BP_OCCLUSION_SENSOR_ERROR, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_LOAD_CELL_FPGA_READ_ERROR, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_DIALYSATE_FLOW_DATA_NOT_RECEIVE, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_CONDUCTIVITY_SENSOR_BAD_STATUS, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID____AVAILABLE_15, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID____AVAILABLE_16, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_CONDUCTIVITY_SENSOR_BAD_STATUS, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_OUTLET_PRIMARY_CONDUCTIVITY_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_PRESSURE_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID____AVAILABLE_17, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, FALSE, FALSE, TRUE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_INLET_WATER_TEMPERATURE_IN_HIGH_RANGE, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, FALSE, FALSE, TRUE , TRUE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE, FALSE, ALARM_ID_FILL_CONDUCTIVITY_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_MEDIUM, 799, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , ALARM_ID_INLET_WATER_TEMPERATURE_IN_HIGH_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE, TRUE , ALARM_ID_FILL_CONDUCTIVITY_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_BATTERY_COMM_FAULT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_SYRINGE_PUMP_STALL, 0, ALARM_ID_NO_ALARM }, // Priority Rank Fault DGFault Stops ClrIm NoClr NoRes NoRin NoEnd NoRcr ClrOnly TxLog ID Escalate In Escalate To { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_NO_CART_SELF_TEST_TIMEOUT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_DRY_SELF_TEST_TIMEOUT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_RTC_COMM_ERROR, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE, TRUE, TRUE, TRUE, TRUE , FALSE, FALSE, ALARM_ID_HD_NVDATAMGMT_CAL_GROUP_RECORD_CRC_INVALID, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, TRUE, ALARM_ID_AIR_TRAP_ILLEGAL_LEVELS, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_AIR_TRAP_ILLEGAL_LEVELS, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_NVDATAMGMT_CAL_GROUP_RECORD_CRC_INVALID, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, FALSE, FALSE, TRUE , TRUE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_INLET_WATER_HIGH_CONDUCTIVITY, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, FALSE, TRUE , TRUE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , ALARM_ID_INLET_WATER_HIGH_CONDUCTIVITY, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_RESTARTED_FAULT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_SYRINGE_PUMP_ADC_ERROR, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_SYRINGE_PUMP_VOLUME_ERROR, 0, ALARM_ID_NO_ALARM }, @@ -561,15 +561,15 @@ { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, FALSE, FALSE, ALARM_ID_HD_BLOOD_LEAK_FAULT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_ARTERIAL_BUBBLE_SELF_TEST_FAILURE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_VENOUS_BUBBLE_SELF_TEST_FAILURE, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_TEMPERATURE_SENSOR_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_TEMPERATURE_SENSOR_ADC_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, TRUE, FALSE, TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, FALSE, ALARM_ID_DG_PRIMARY_HEATER_INTERNAL_TEMP_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, TRUE, FALSE, TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, FALSE, ALARM_ID_DG_PRIMARY_HEATER_CJ_TEMP_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, TRUE, FALSE, TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, FALSE, ALARM_ID_DG_TRIMMER_HEATER_INTERNAL_TEMP_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, TRUE, FALSE, TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, FALSE, ALARM_ID_DG_TRIMMER_HEATER_CJ_TEMP_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, TRUE, FALSE, TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, FALSE, FALSE, ALARM_ID_DG_MAIN_PRIMARY_HEATER_VOLTAGE_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, TRUE, FALSE, TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, FALSE, FALSE, ALARM_ID_DG_SMALL_PRIMARY_HEATER_VOLTAGE_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, TRUE, FALSE, TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, FALSE, FALSE, ALARM_ID_DG_TRIMMER_HEATER_VOLTAGE_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_TEMPERATURE_SENSOR_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_TEMPERATURE_SENSOR_ADC_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE, FALSE, TRUE , TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, FALSE, ALARM_ID____AVAILABLE_33, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE, FALSE, TRUE , TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, FALSE, ALARM_ID____AVAILABLE_32, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE, FALSE, TRUE , TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, FALSE, ALARM_ID____AVAILABLE_31, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE, FALSE, TRUE , TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, FALSE, ALARM_ID____AVAILABLE_30, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, FALSE, FALSE, ALARM_ID_DG_MAIN_PRIMARY_HEATER_VOLTAGE_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, FALSE, FALSE, ALARM_ID_DG_SMALL_PRIMARY_HEATER_VOLTAGE_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, FALSE, FALSE, ALARM_ID_DG_TRIMMER_HEATER_VOLTAGE_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 4, FALSE, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, ALARM_ID_END_OF_TREATMENT_HIGH, 0, ALARM_ID_NO_ALARM }, // Priority Rank Fault DGFault Stops ClrIm NoClr NoRes NoRin NoEnd NoRcr ClrOnly TxLog ID Escalate In Escalate To { ALARM_PRIORITY_HIGH, 5, FALSE, FALSE, TRUE , TRUE , FALSE, TRUE, TRUE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_TREATMENT_STOPPED_NO_RINSEBACK, 0, ALARM_ID_NO_ALARM }, @@ -580,7 +580,7 @@ { ALARM_PRIORITY_HIGH, 11, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE, ALARM_ID_VENOUS_PRESSURE_HIGH, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 12, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE, ALARM_ID_ARTERIAL_PRESSURE_LOW, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 13, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE, ALARM_ID_ARTERIAL_PRESSURE_HIGH, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 14, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE, ALARM_ID_DG_FLUID_LEAK_DETECTED, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 14, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE, ALARM_ID_DG_FLUID_LEAK_DETECTED, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 15, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE, ALARM_ID_HD_FLUID_LEAK_DETECTED, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 16, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_HD_SHOCK, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 16, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_SHOCK, 0, ALARM_ID_NO_ALARM }, @@ -590,23 +590,23 @@ { ALARM_PRIORITY_HIGH, 19, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE, ALARM_ID_DG_COMM_TIMEOUT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 20, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE, ALARM_ID_AIR_TRAP_FILL_DURING_TREATMENT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 21, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE, ALARM_ID_OCCLUSION_BLOOD_PUMP, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 21, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, TRUE, ALARM_ID_DG_DIALYSATE_TEMPERATURE_SENSORS_DRIFT_TIMEOUT, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_LOW, 21, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE, ALARM_ID_DG_SW_CONFIG_RECORD_INVALID_CRC, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE, ALARM_ID_ACID_CONDUCTIVITY_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 23, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE, ALARM_ID_DIALYSATE_CONDUCTIVITY_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 24, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE, TRUE, FALSE, ALARM_ID_CREATING_DIALYSATE_PLEASE_WAIT, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE, TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, FALSE, TRUE, ALARM_ID_INLET_WATER_HIGH_TEMPERATURE, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 28, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE, ALARM_ID____AVAILABLE_28, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_DIALYSATE_TEMPERATURE_SENSORS_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_LOW, 21, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_DG_SW_CONFIG_RECORD_INVALID_CRC, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_ACID_CONDUCTIVITY_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 23, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID____AVAILABLE_34, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 24, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE, TRUE, FALSE, ALARM_ID_CREATING_DIALYSATE_PLEASE_WAIT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE, TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, FALSE, FALSE, ALARM_ID_INLET_WATER_HIGH_TEMPERATURE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 28, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID____AVAILABLE_28, 0, ALARM_ID_NO_ALARM }, // Priority Rank Fault DGFault Stops ClrIm NoClr NoRes NoRin NoEnd NoRcr ClrOnly TxLog ID Escalate In Escalate To - { ALARM_PRIORITY_LOW, 30, FALSE, FALSE, FALSE, TRUE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_INLET_WATER_CONDUCTIVITY_IN_HIGH_RANGE, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 32, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE, ALARM_ID_INLET_WATER_CONDUCTIVITY_IN_LOW_RANGE, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, FALSE, FALSE, TRUE , FALSE, FALSE, TRUE, FALSE, FALSE, FALSE, FALSE, TRUE, ALARM_ID_INLET_WATER_PRESSURE_IN_LOW_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_MEDIUM, 799, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , ALARM_ID_INLET_WATER_CONDUCTIVITY_IN_HIGH_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_MEDIUM, 799, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , ALARM_ID_INLET_WATER_CONDUCTIVITY_IN_LOW_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_MEDIUM, 799, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , ALARM_ID_INLET_WATER_PRESSURE_IN_LOW_RANGE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 36, FALSE, FALSE, FALSE, TRUE, FALSE, TRUE, TRUE , FALSE, FALSE, FALSE, FALSE, ALARM_ID_PRIME_COMPLETED_HIGH, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_NVDATA_EEPROM_OPS_FAILURE, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_LOW, 799, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE, ALARM_ID_HD_SW_CONFIG_RECORD_INVALID_CRC, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_LOW, 799, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_HD_SW_CONFIG_RECORD_INVALID_CRC, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_NVDATA_HW_USAGE_DATA_CRC_ERROR, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_CHEM_DISINFECT_PRIME_ACID_LINE_TIME_OUT, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, FALSE, FALSE, TRUE, FALSE, TRUE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_INLET_WATER_TEMPERATURE_IN_LOW_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_MEDIUM, 799, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , ALARM_ID_INLET_WATER_TEMPERATURE_IN_LOW_RANGE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_DIAL_IN_PUMP_FLOW_VS_MOTOR_SPEED_CHECK, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_BLOOD_PUMP_ROTOR_SPEED_TOO_HIGH, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID____AVAILABLE_26, 0, ALARM_ID_NO_ALARM }, @@ -615,11 +615,11 @@ { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_HD_LOAD_CELL_ACCELERATION_RES_2_ALARM, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , TRUE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, ALARM_ID_TREATMENT_RINSEBACK_TIMEOUT_ALARM, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , TRUE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, ALARM_ID_TREATMENT_RECIRC_TIMEOUT_ALARM, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_CARTRIDGE_DOOR_OPENED, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_CARTRIDGE_DOOR_OPENED, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_HD_ACTIVE_RESERVOIR_DEPLETION_TIME_OUT, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DIALYSATE_FLOW_RATE_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DIALYSATE_FLOW_RATE_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_HD_SYRINGE_PUMP_SYRINGE_EMPTY, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE, ALARM_ID_HD_SYRINGE_PUMP_OCCLUSION, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_HD_SYRINGE_PUMP_OCCLUSION, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, ALARM_ID_HD_SYRINGE_PUMP_NOT_ENOUGH_HEPARIN_ALARM, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , TRUE, TRUE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE, ALARM_ID_HD_ARTERIAL_BUBBLE_DETECTED, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , TRUE, TRUE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE, ALARM_ID_HD_ARTERIAL_BUBBLE_DETECTED_RINSEBACK, 0, ALARM_ID_NO_ALARM }, @@ -628,37 +628,37 @@ { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_HD_RESERVOIR_FULL_AND_DG_NOT_READY_TO_SWITCH, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_SOFTWARE_FAULT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_HD_COMM_TIMEOUT, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 799, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_FPGA_COMM_TIMEOUT, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 799, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_LOAD_CELL_ADC_ERROR, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 799, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_LOAD_CELLS_TARE_WEIGHT_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_FPGA_COMM_TIMEOUT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_LOAD_CELL_ADC_ERROR, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 799, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_LOAD_CELLS_TARE_WEIGHT_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_LOAD_CELLS_INVALID_CALIBRATION, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_INVALID_LOAD_CELL_VALUE, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 799, FALSE, TRUE, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_UV_REACTOR_NOT_HEALTHY, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_UV_REACTOR_NOT_HEALTHY, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, FALSE, TRUE, FALSE, FALSE, TRUE , TRUE , FALSE, TRUE , FALSE, ALARM_ID_DG_FAN_RPM_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_CONCENTRATE_PUMP_FAULT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_CP1_SPEED_CONTROL_ERROR, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_CP2_SPEED_CONTROL_ERROR, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DRAIN_PUMP_RPM_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 799, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DRAIN_PUMP_OFF_FAULT, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_FLOW_RATE_OUT_OF_UPPER_RANGE, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_FLOW_RATE_OUT_OF_LOWER_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DRAIN_PUMP_RPM_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DRAIN_PUMP_OFF_FAULT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_FLOW_RATE_OUT_OF_UPPER_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_FLOW_RATE_OUT_OF_LOWER_RANGE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_RO_PUMP_FLOW_RATE_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_RO_PUMP_DUTY_CYCLE_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_RO_PUMP_DUTY_CYCLE_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_RO_PUMP_PRESSURE_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 799, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_TEMPERATURE_SENSOR_FAULT, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 799, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_TEMPERATURE_SENSORS_ADC_FAULT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_TEMPERATURE_SENSOR_FAULT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_TEMPERATURE_SENSORS_ADC_FAULT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_HEATERS_NEGATIVE_COLD_JUNCTION_TEMPERATURE, 0, ALARM_ID_NO_ALARM }, // Priority Rank Fault DGFault Stops ClrIm NoClr NoRes NoRin NoEnd NoRcr ClrOnly TxLog ID Escalate In Escalate To { ALARM_PRIORITY_HIGH, 799, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_HEATERS_FAULT, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_THERMISTORS_TEMPERATURE_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_DG_BAD_INLET_WATER_QUALITY, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 799, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_INLET_WATER_PRESSURE_FAULT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_THERMISTORS_TEMPERATURE_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID____AVAILABLE_29, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 799, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID____AVAILABLE_35, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_PRESSURE_SENSOR_FAULT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_RO_REJECTION_RATIO_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 799, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_CONDUCTIVITY_SENSOR_FAULT, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_DIALYSATE_FILL_OUT_OF_TIME, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_CONDUCTIVITY_SENSOR_FAULT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE, TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_DIALYSATE_FILL_OUT_OF_TIME, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_FLOW_METER_CHECK_FAILURE, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 799, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_PRIMARY_HEATER_ON_WITH_NO_FLOW_TIMEOUT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_PRIMARY_HEATER_ON_WITH_NO_FLOW_TIMEOUT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_DRAIN_CIRCULATION_LINE_TIMEOUT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_HD_BATTERY_PACK_ERROR_DETECTED, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_MEDIUM, 800, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_BLOOD_SITTING_WARNING, 0, ALARM_ID_NO_ALARM }, @@ -668,12 +668,12 @@ { ALARM_PRIORITY_MEDIUM, 899, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_RINSEBACK_VOLUME_CHECK_FAILURE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_MEDIUM, 899, FALSE, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, ALARM_ID_END_TREATMENT_TIMEOUT_ALARM, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_MEDIUM, 899, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_BLOOD_PRIME_VOLUME_CHECK_FAILURE, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_MEDIUM, 899, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_HD_SYRINGE_DETECTED, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_MEDIUM, 899, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_HD_SYRINGE_DETECTED, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH , 799, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_HD_SYRINGE_PUMP_SYRINGE_REMOVED, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_MEDIUM, 899, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_HD_BATTERY_PACK_CHARGE_TOO_LOW, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_LOW, 900, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_EMPTY_SALINE_BAG, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_LOW, 901, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_DIALYSATE_TEMPERATURE_HIGH, ALM_ESC_5_MIN, ALARM_ID_TREATMENT_STOPPED_NO_RINSEBACK }, - { ALARM_PRIORITY_LOW, 902, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_DIALYSATE_TEMPERATURE_LOW, ALM_ESC_5_MIN, ALARM_ID_TREATMENT_STOPPED_NO_RINSEBACK }, + { ALARM_PRIORITY_LOW, 901, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , ALARM_ID_DIALYSATE_TEMPERATURE_HIGH, ALM_ESC_5_MIN, ALARM_ID_TREATMENT_STOPPED_NO_RINSEBACK }, + { ALARM_PRIORITY_LOW, 902, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , ALARM_ID_DIALYSATE_TEMPERATURE_LOW, ALM_ESC_5_MIN, ALARM_ID_TREATMENT_STOPPED_NO_RINSEBACK }, { ALARM_PRIORITY_LOW, 903, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_TREATMENT_STOPPED_BY_USER, 0, ALARM_ID_NO_ALARM }, // Priority Rank Fault DGFault Stops ClrIm NoClr NoRes NoRin NoEnd NoRcr ClrOnly TxLog ID Escalate In Escalate To { ALARM_PRIORITY_LOW, 904, FALSE, FALSE, FALSE, TRUE , FALSE, TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_END_OF_TREATMENT_WARNING, 0, ALARM_ID_NO_ALARM }, @@ -690,8 +690,8 @@ { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_INSTALL_NEW_CARTRIDGE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, ALARM_ID_PRIME_SALINE_DIALYZER_TIME_OUT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_NO_CARTRIDGE_LOADED, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, FALSE, TRUE , FALSE, TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_CARTRIDGE_REMOVAL_FAILURE, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE, ALARM_ID_BICARB_CONDUCTIVITY_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_CARTRIDGE_REMOVAL_FAILURE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_BICARB_CONDUCTIVITY_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_RESERVOIR_DRAIN_TIMEOUT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_RESERVOIR_FILL_TIMEOUT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_RESERVOIR_LEAK_TIMEOUT, 0, ALARM_ID_NO_ALARM }, @@ -709,7 +709,7 @@ { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_INVALID_SERVICE_RECORD_CRC, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_UI_COMPATIBILITY_ERROR, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_DG_COMPATIBILITY_ERROR, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_FPGA_POWER_OUT_TIMEOUT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_FPGA_POWER_OUT_TIMEOUT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_FPGA_POWER_OUT_TIMEOUT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_TEMPERATURES_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_UI_POST_FAILURE_SHASUM, 0, ALARM_ID_NO_ALARM }, @@ -734,14 +734,14 @@ { ALARM_PRIORITY_HIGH, 3, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_HD_ACTIVE_RESERVOIR_WEIGHT_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_DIALYSATE_DRAIN_TIME_OUT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_ARTERIAL_PRESSURE_READ_TIMEOUT_ERROR, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, TRUE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_DG_ACID_BOTTLE_LOW_VOLUME, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, TRUE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_DG_BICARB_BOTTLE_LOW_VOLUME, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, TRUE, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_LOAD_CELL_WEIGHT_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_DG_ACID_BOTTLE_LOW_VOLUME, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_DG_BICARB_BOTTLE_LOW_VOLUME, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_LOAD_CELL_WEIGHT_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_LOAD_CELL_PRIMARY_BACKUP_DRIFT_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_MEDIUM, 899, FALSE, FALSE, FALSE, TRUE , FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, ALARM_ID_HD_TREATMENT_RECIRC_TIMEOUT_WARNING, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_MEDIUM, 899, FALSE, FALSE, FALSE, TRUE , FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, ALARM_ID_HD_TREATMENT_RINSEBACK_TIMEOUT_WARNING, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_FPGA_CLOCK_SPEED_CHECK_FAILURE, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 799, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_TRIMMER_HEATER_ON_WITH_NO_FLOW_TIMEOUT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_TRIMMER_HEATER_ON_WITH_NO_FLOW_TIMEOUT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_DIALYSATE_OR_CONC_CAP_NOT_IN_PROPER_POSITION, 0, ALARM_ID_NO_ALARM }, }; // Priority Rank Fault DGFault Stops ClrIm NoClr NoRes NoRin NoEnd NoRcr ClrOnly TxLog ID Escalate In Escalate To @@ -848,13 +848,12 @@ { 3, ALARM_ID_DG_HEATING_INVALID_CAL_RECORD }, { 3, ALARM_ID_DG_CONC_PUMP_HALL_SENSOR_OUT_OF_RANGE }, { 3, ALARM_ID_HD_BP_OCCLUSION_SENSOR_ERROR }, - { 3, ALARM_ID____AVAILABLE_25 }, + { 3, ALARM_ID_DG_LOAD_CELL_FPGA_READ_ERROR }, { 3, ALARM_ID_HD_DIALYSATE_FLOW_DATA_NOT_RECEIVE }, { 3, ALARM_ID_DG_CONDUCTIVITY_SENSOR_BAD_STATUS }, - { 3, ALARM_ID____AVAILABLE_15 }, - { 3, ALARM_ID____AVAILABLE_16 }, + { 3, ALARM_ID_DG_OUTLET_PRIMARY_CONDUCTIVITY_OUT_OF_RANGE }, + { 3, ALARM_ID_DG_PRESSURE_OUT_OF_RANGE }, { 3, ALARM_ID____AVAILABLE_17 }, - { 3, ALARM_ID_INLET_WATER_TEMPERATURE_IN_HIGH_RANGE }, { 3, ALARM_ID_FILL_CONDUCTIVITY_OUT_OF_RANGE }, { 3, ALARM_ID_HD_BATTERY_COMM_FAULT }, { 3, ALARM_ID_HD_SYRINGE_PUMP_STALL }, @@ -875,10 +874,10 @@ { 3, ALARM_ID_HD_VENOUS_BUBBLE_SELF_TEST_FAILURE }, { 3, ALARM_ID_DG_TEMPERATURE_SENSOR_OUT_OF_RANGE }, { 3, ALARM_ID_DG_TEMPERATURE_SENSOR_ADC_OUT_OF_RANGE }, - { 3, ALARM_ID_DG_PRIMARY_HEATER_INTERNAL_TEMP_OUT_OF_RANGE }, - { 3, ALARM_ID_DG_PRIMARY_HEATER_CJ_TEMP_OUT_OF_RANGE }, - { 3, ALARM_ID_DG_TRIMMER_HEATER_INTERNAL_TEMP_OUT_OF_RANGE }, - { 3, ALARM_ID_DG_TRIMMER_HEATER_CJ_TEMP_OUT_OF_RANGE }, + { 3, ALARM_ID____AVAILABLE_33 }, + { 3, ALARM_ID____AVAILABLE_32 }, + { 3, ALARM_ID____AVAILABLE_31 }, + { 3, ALARM_ID____AVAILABLE_30 }, { 3, ALARM_ID_DG_MAIN_PRIMARY_HEATER_VOLTAGE_OUT_OF_RANGE }, { 3, ALARM_ID_DG_SMALL_PRIMARY_HEATER_VOLTAGE_OUT_OF_RANGE }, { 3, ALARM_ID_DG_TRIMMER_HEATER_VOLTAGE_OUT_OF_RANGE }, @@ -911,9 +910,22 @@ { 3, ALARM_ID_BICARB_CONDUCTIVITY_OUT_OF_RANGE }, { 3, ALARM_ID_DG_FLOW_METER_CHECK_FAILURE }, { 3, ALARM_ID_INLET_WATER_HIGH_TEMPERATURE }, - { 3, ALARM_ID_INLET_WATER_TEMPERATURE_IN_LOW_RANGE }, - { 3, ALARM_ID_INLET_WATER_PRESSURE_IN_LOW_RANGE }, { 3, ALARM_ID_RO_REJECTION_RATIO_OUT_OF_RANGE }, + { 3, ALARM_ID_DG_DIALYSATE_FILL_OUT_OF_TIME }, + { 3, ALARM_ID_DG_DIALYSATE_TEMPERATURE_SENSORS_OUT_OF_RANGE }, + { 3, ALARM_ID_DIALYSATE_FLOW_RATE_OUT_OF_RANGE }, + { 3, ALARM_ID_FLOW_RATE_OUT_OF_UPPER_RANGE }, + { 3, ALARM_ID_FLOW_RATE_OUT_OF_LOWER_RANGE }, + { 3, ALARM_ID_DRAIN_PUMP_RPM_OUT_OF_RANGE }, + { 3, ALARM_ID_DRAIN_PUMP_OFF_FAULT }, + { 3, ALARM_ID_UV_REACTOR_NOT_HEALTHY }, + { 3, ALARM_ID_DG_TEMPERATURE_SENSORS_ADC_FAULT }, + { 3, ALARM_ID_DG_TEMPERATURE_SENSOR_FAULT }, + { 3, ALARM_ID_DG_LOAD_CELL_ADC_ERROR }, + { 3, ALARM_ID_DG_FPGA_COMM_TIMEOUT }, + { 3, ALARM_ID_DG_PRIMARY_HEATER_ON_WITH_NO_FLOW_TIMEOUT }, + { 3, ALARM_ID_DG_TRIMMER_HEATER_ON_WITH_NO_FLOW_TIMEOUT }, + { 3, ALARM_ID_CONDUCTIVITY_SENSOR_FAULT }, { 4, ALARM_ID_END_OF_TREATMENT_HIGH }, { 5, ALARM_ID_TREATMENT_STOPPED_NO_RINSEBACK }, { 6, ALARM_ID_HD_BLOOD_LEAK_DETECTED }, @@ -933,13 +945,10 @@ { 19, ALARM_ID_DG_COMM_TIMEOUT }, { 20, ALARM_ID_AIR_TRAP_FILL_DURING_TREATMENT }, { 21, ALARM_ID_OCCLUSION_BLOOD_PUMP }, - { 21, ALARM_ID_DG_DIALYSATE_TEMPERATURE_SENSORS_DRIFT_TIMEOUT }, { 21, ALARM_ID_DG_SW_CONFIG_RECORD_INVALID_CRC }, - { 23, ALARM_ID_DIALYSATE_CONDUCTIVITY_OUT_OF_RANGE }, + { 23, ALARM_ID____AVAILABLE_34 }, { 24, ALARM_ID_CREATING_DIALYSATE_PLEASE_WAIT }, { 28, ALARM_ID____AVAILABLE_28 }, - { 30, ALARM_ID_INLET_WATER_CONDUCTIVITY_IN_HIGH_RANGE }, - { 32, ALARM_ID_INLET_WATER_CONDUCTIVITY_IN_LOW_RANGE }, { 36, ALARM_ID_PRIME_COMPLETED_HIGH }, { 799, ALARM_ID_NVDATA_EEPROM_OPS_FAILURE }, { 799, ALARM_ID_HD_SW_CONFIG_RECORD_INVALID_CRC }, @@ -955,7 +964,6 @@ { 799, ALARM_ID_PUMP_TRACK_LATCH_OPENED }, { 799, ALARM_ID_CARTRIDGE_DOOR_OPENED }, { 799, ALARM_ID_HD_ACTIVE_RESERVOIR_DEPLETION_TIME_OUT }, - { 799, ALARM_ID_DIALYSATE_FLOW_RATE_OUT_OF_RANGE }, { 799, ALARM_ID_HD_SYRINGE_PUMP_SYRINGE_EMPTY }, { 799, ALARM_ID_HD_SYRINGE_PUMP_OCCLUSION }, { 799, ALARM_ID_HD_SYRINGE_PUMP_NOT_ENOUGH_HEPARIN_ALARM }, @@ -965,36 +973,28 @@ { 799, ALARM_ID_HD_RESERVOIR_FULL_AND_DG_NOT_READY_TO_SWITCH }, { 799, ALARM_ID_DG_SOFTWARE_FAULT }, { 799, ALARM_ID_HD_COMM_TIMEOUT }, - { 799, ALARM_ID_DG_FPGA_COMM_TIMEOUT }, - { 799, ALARM_ID_DG_LOAD_CELL_ADC_ERROR }, { 799, ALARM_ID_DG_LOAD_CELLS_TARE_WEIGHT_OUT_OF_RANGE }, { 799, ALARM_ID_DG_LOAD_CELLS_INVALID_CALIBRATION }, - { 799, ALARM_ID_UV_REACTOR_NOT_HEALTHY }, { 799, ALARM_ID_DG_CONCENTRATE_PUMP_FAULT }, { 799, ALARM_ID_CP1_SPEED_CONTROL_ERROR }, { 799, ALARM_ID_CP2_SPEED_CONTROL_ERROR }, - { 799, ALARM_ID_DRAIN_PUMP_RPM_OUT_OF_RANGE }, - { 799, ALARM_ID_DRAIN_PUMP_OFF_FAULT }, - { 799, ALARM_ID_FLOW_RATE_OUT_OF_UPPER_RANGE }, - { 799, ALARM_ID_FLOW_RATE_OUT_OF_LOWER_RANGE }, { 799, ALARM_ID_RO_PUMP_FLOW_RATE_OUT_OF_RANGE }, { 799, ALARM_ID_RO_PUMP_PRESSURE_OUT_OF_RANGE }, - { 799, ALARM_ID_DG_TEMPERATURE_SENSOR_FAULT }, - { 799, ALARM_ID_DG_TEMPERATURE_SENSORS_ADC_FAULT }, { 799, ALARM_ID_DG_HEATERS_NEGATIVE_COLD_JUNCTION_TEMPERATURE }, { 799, ALARM_ID_DG_HEATERS_FAULT }, - { 799, ALARM_ID_DG_BAD_INLET_WATER_QUALITY }, - { 799, ALARM_ID_INLET_WATER_PRESSURE_FAULT }, + { 799, ALARM_ID____AVAILABLE_29 }, + { 799, ALARM_ID____AVAILABLE_35 }, { 799, ALARM_ID_PRESSURE_SENSOR_FAULT }, - { 799, ALARM_ID_CONDUCTIVITY_SENSOR_FAULT }, - { 799, ALARM_ID_DG_DIALYSATE_FILL_OUT_OF_TIME }, - { 799, ALARM_ID_DG_PRIMARY_HEATER_ON_WITH_NO_FLOW_TIMEOUT }, { 799, ALARM_ID_DG_DRAIN_CIRCULATION_LINE_TIMEOUT }, { 799, ALARM_ID_HD_BATTERY_PACK_ERROR_DETECTED }, - { 799, ALARM_ID_DG_TRIMMER_HEATER_ON_WITH_NO_FLOW_TIMEOUT }, { 799, ALARM_ID_DG_RESERVOIR_LEAK_TIMEOUT }, { 799, ALARM_ID_HD_SYRINGE_PUMP_SYRINGE_REMOVED }, { 799, ALARM_ID_DG_DIALYSATE_OR_CONC_CAP_NOT_IN_PROPER_POSITION }, + { 799, ALARM_ID_INLET_WATER_CONDUCTIVITY_IN_HIGH_RANGE }, + { 799, ALARM_ID_INLET_WATER_CONDUCTIVITY_IN_LOW_RANGE }, + { 799, ALARM_ID_INLET_WATER_TEMPERATURE_IN_LOW_RANGE }, + { 799, ALARM_ID_INLET_WATER_TEMPERATURE_IN_HIGH_RANGE }, + { 799, ALARM_ID_INLET_WATER_PRESSURE_IN_LOW_RANGE }, { 800, ALARM_ID_BLOOD_SITTING_WARNING }, { 801, ALARM_ID_END_OF_TREATMENT_ALARM }, { 802, ALARM_ID_PRIME_COMPLETED_MEDIUM },