Index: AlarmDefs.h =================================================================== diff -u -r4c26156e96966916dba4e89cdb355c06f0b19d19 -r2ae4b6cd38654b8372d3742b8f53be8651aca4e4 --- AlarmDefs.h (.../AlarmDefs.h) (revision 4c26156e96966916dba4e89cdb355c06f0b19d19) +++ AlarmDefs.h (.../AlarmDefs.h) (revision 2ae4b6cd38654b8372d3742b8f53be8651aca4e4) @@ -7,8 +7,8 @@ * * @file AlarmDefs.h * -* @author (last) Vinayakam Mani -* @date (last) 05-Sep-2023 +* @author (last) Dara Navaei +* @date (last) 11-Sep-2023 * * @author (original) Sean * @date (original) 11-Feb-2020 @@ -1227,7 +1227,7 @@ { 60 , ALARM_ID_HD_VALVE_NOT_FUNCTIONAL , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Valve Control | HD valve not functional | If the valve bit masks are not as expected in different valve control mode. | "}, { 61 , ALARM_ID_HD_VALVE_CURRENT_OUT_OF_RANGE , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Valve Current | HD valve current out of range | If the valve current is over the threshold of 1.0 A for 1000 msec. | "}, { 62 , ALARM_ID_HD_VALVE_POSITION_OUT_OF_RANGE , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Valve Position | HD valve position out of target | If the current position has deviated from the expected position for more than 1 second. | "}, - { 63 , ALARM_ID_HD_SYRINGE_PUMP_PRIME_TIMEOUT , "HD | Heparin Syringe Pump Prime Timeout | A problem has been detected with the heparin pump.\n- Ensure heparin line is not kinked or clamped. | Process: Pre- Tx Hep. Pump Timeout. | HD syringe pump timeout alarm | If the syringe pump priming > 5 seconds | "}, + { 63 , ALARM_ID_HD_SYRINGE_PUMP_PRIME_TIMEOUT , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Pre- Tx Hep. Pump Timeout. | HD syringe pump timeout alarm | If the syringe pump priming > 5 seconds | "}, { 64 , ALARM_ID_DG_BARO_PRESSURE_OUT_OF_RANGE , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: Baro Range | Barometric pressure is out of range | If the barometric pressure is > 10.1 psi or < 15.4 psi for a certain period of time. | "}, { 65 , ALARM_ID_HD_DG_COMMAND_INVALID_PARAMETER_FAULT , "HD | Communication Error | A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | Loss of DG Comm (Reject) | HD requests DG command with invalid parameter fault | If the latest DG command response indicated invalid parameter | "}, { 66 , ALARM_ID_HD_BLOOD_LEAK_SENSOR_SET_POINT_SET_FAILURE , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: BLD Config Set Point | HD blood leak sensor set point set failure | If the set point of the blood leak sensor does not match the set point in the calibration record. | "}, @@ -1422,9 +1422,9 @@ { 255 , ALARM_ID_HD_UI_COMPATIBILITY_ERROR , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: UI Compatibility | HD and UI software builds are not compatible | If the compatibility revision field of the UI is not equal to SW_COMPATIBILITY_REV. | "}, { 256 , ALARM_ID_HD_DISINFECT_CHEM_FLUSH , "HD | Chemical Flush in Progress | Chemical flush in progress.\n- Please wait for the process to complete. \n- You will not be able to create a treatment until the process\nis finished.\n- To end this process, press STOP on the front of the device. | Cleaning: Chem Flush In Progress | DG flush started | If DG flush successfully started from the UI. | "}, { 257 , ALARM_ID_AVAILABLE_8 , " | | | | | | "}, - { 258 , ALARM_ID_HD_UI_POST_FAILURE_OS_VERSION , "UI HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: OS Version | UI POST OS Version compatibility failure | If the UI POST reports the OS Version compatibility faillure. | "}, + { 258 , ALARM_ID_HD_UI_POST_FAILURE_OS_VERSION , "UI HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: OS Version | UI POST OS Version compatibility failure | If the UI POST reports the OS Version compatibility failure. | "}, { 259 , ALARM_ID_HD_TEMPERATURES_OUT_OF_RANGE , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Temp Sensors Error | HD temperatures out of range | If thermistor reports temperature > 80.0 C or < 0.0 C for a certain period of time. | "}, - { 260 , ALARM_ID_HD_UI_POST_FAILURE_SHASUM , "UI HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: UI Image | UI POST Application Integrity (Sha256Sum) failure | If the UI POST reports a shasum faillure. | "}, + { 260 , ALARM_ID_HD_UI_POST_FAILURE_SHASUM , "UI HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: UI Image | UI POST Application Integrity (Sha256Sum) failure | If the UI POST reports a shasum failure. | "}, { 261 , ALARM_ID_HD_UI_POST_FAILURE_CANBUS , "UI HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: UI CAN Driver | UI POST CANBus failure | If the UI POST reports a can bus failure. | "}, { 262 , ALARM_ID_HD_UI_POST_FAILURE_DISPLAY , " | | | | UI POST Display failure | If the UI POST reports a display failure. | "}, { 263 , ALARM_ID_HD_UI_POST_FAILURE_TOUCH , "UI HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: UI Touch Driver | UI POST Touch failure | If the UI POST reports a touch failure. | "},