Index: AlarmDefs.h =================================================================== diff -u -rb9965ff34817dcd69c73c588c634e6d8abd3a4c3 -r13d6de30c57052b0422a05f3d3ea54385888fa6b --- AlarmDefs.h (.../AlarmDefs.h) (revision b9965ff34817dcd69c73c588c634e6d8abd3a4c3) +++ AlarmDefs.h (.../AlarmDefs.h) (revision 13d6de30c57052b0422a05f3d3ea54385888fa6b) @@ -120,7 +120,7 @@ ALARM_ID_HD_SYRINGE_PUMP_DAC_WRITE_ERROR = 82, ///< HD syringe pump DAC write failure ALARM_ID_HD_SYRINGE_PUMP_RUNNING_WHILE_BP_OFF_ERROR = 83, ///< HD syringe pump is running while the BP is off ALARM_ID_HD_VENOUS_PRESSURE_READ_TIMEOUT_ERROR = 84, ///< HD venous pressure sensor not being read - ALARM_ID_____AVAILABLE_25 = 85, ///< This alarm ID is available for use + ALARM_ID_PUMP_TRACK_LATCH_OPENED = 85, ///< HD pump track latch opened alarm ALARM_ID_HD_BP_OCCLUSION_READ_TIMEOUT_ERROR = 86, ///< HD BP occlusion sensor not being read ALARM_ID_DG_HEATING_INVALID_CAL_RECORD = 87, ///< DG heating invalid calibration record ALARM_ID_DG_CONC_PUMP_HALL_SENSOR_OUT_OF_RANGE = 88, ///< DG concentrate pumps hall sensor out of range @@ -530,7 +530,7 @@ { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_SYRINGE_PUMP_DAC_WRITE_ERROR, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_SYRINGE_PUMP_RUNNING_WHILE_BP_OFF_ERROR, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_VENOUS_PRESSURE_READ_TIMEOUT_ERROR, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_____AVAILABLE_25, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , FALSE, FALSE, TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, ALARM_ID_PUMP_TRACK_LATCH_OPENED, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_BP_OCCLUSION_READ_TIMEOUT_ERROR, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, FALSE, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_HEATING_INVALID_CAL_RECORD, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , TRUE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_CONC_PUMP_HALL_SENSOR_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, @@ -843,7 +843,6 @@ { 3, ALARM_ID_HD_SYRINGE_PUMP_DAC_WRITE_ERROR }, { 3, ALARM_ID_HD_SYRINGE_PUMP_RUNNING_WHILE_BP_OFF_ERROR }, { 3, ALARM_ID_HD_VENOUS_PRESSURE_READ_TIMEOUT_ERROR }, - { 3, ALARM_ID_____AVAILABLE_25 }, { 3, ALARM_ID_HD_BP_OCCLUSION_READ_TIMEOUT_ERROR }, { 3, ALARM_ID_DG_HEATING_INVALID_CAL_RECORD }, { 3, ALARM_ID_DG_CONC_PUMP_HALL_SENSOR_OUT_OF_RANGE }, @@ -943,6 +942,7 @@ { 799, ALARM_ID_HD_LOAD_CELL_ACCELERATION_RES_2_ALARM }, { 799, ALARM_ID_TREATMENT_RINSEBACK_TIMEOUT_ALARM }, { 799, ALARM_ID_TREATMENT_RECIRC_TIMEOUT_ALARM }, + { 799, ALARM_ID_PUMP_TRACK_LATCH_OPENED }, { 799, ALARM_ID_CARTRIDGE_DOOR_OPENED }, { 799, ALARM_ID_HD_ACTIVE_RESERVOIR_DEPLETION_TIME_OUT }, { 799, ALARM_ID_DIALYSATE_FLOW_RATE_OUT_OF_RANGE },