Index: AlarmDefs.h =================================================================== diff -u -rc9c2dec290efbbb4c1a61b7029c774111d1007e2 -re18ec8e1d17b3e29993c966c53992b18cda5a799 --- AlarmDefs.h (.../AlarmDefs.h) (revision c9c2dec290efbbb4c1a61b7029c774111d1007e2) +++ AlarmDefs.h (.../AlarmDefs.h) (revision e18ec8e1d17b3e29993c966c53992b18cda5a799) @@ -60,7 +60,14 @@ ALARM_ID_VENOUS_PRESSURE_HIGH = 35, ///< Venous pressure too high during treatment. ALARM_ID_UF_RATE_ACCURACY_ERROR = 36, ///< Ultrafiltration rate accuracy error during treatment. ALARM_ID_UF_VOLUME_ACCURACY_ERROR = 37, ///< Ultrafiltration volume accuracy error during treatment. - ALARM_ID_NVDATA_MFG_RECORD_CRC_ERROR = 38, ///< Manufacturing record CRC failure. + ALARM_ID_RTC_BATTERY_LOW = 38, ///< RTC battery low. + ALARM_ID_RTC_OR_TIMER_ACCURACY_FAILURE = 39, ///< RTC or MCU timer inaccurate. + ALARM_ID_NVDATA_EEPROM_OPS_FAILURE = 40, ///< EEPRON operations (read, write, erase) failure. + ALARM_ID_RTC_RAM_OPS_ERROR = 41, ///< RTC (i.e read time) or RAM operations failure (read or write). + ALARM_ID_NVDATA_MFG_RECORD_CRC_ERROR = 42, ///< Manufacturing record CRC failure. + ALARM_ID_NVDATA_SRVC_RECORD_CRC_ERROR = 43, ///< Service record CRC failure. + ALARM_ID_NVDATA_CAL_RECORD_CRC_ERROR = 44, ///< Calibration record CRC failure. + ALARM_ID_NVDATA_HW_USAGE_DATA_CRC_ERROR = 45, ///< HW usage data (treatment time in HD and total consumed water in DG) failure. NUM_OF_ALARM_IDS ///< Total # of alarms. } ALARM_ID_T;