Index: AlarmDefs.h =================================================================== diff -u -r9f5af4b529a84a658f814f23564c6d7ee13c5c20 -r3c53c91362c8b14deaa66b1eabe139d71f990d28 --- AlarmDefs.h (.../AlarmDefs.h) (revision 9f5af4b529a84a658f814f23564c6d7ee13c5c20) +++ AlarmDefs.h (.../AlarmDefs.h) (revision 3c53c91362c8b14deaa66b1eabe139d71f990d28) @@ -1459,7 +1459,7 @@ { 298 , ALARM_ID_DG_BARO_SENSOR_FPGA_FAULT , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG Fault: Baro Pres Comm | DG barometric sensor FPGA fault | If the FPGA does not report fresh data for a certain period of time | "}, { 299 , ALARM_ID_DG_INVALID_SERIAL_NUMBER , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG POST: Serial Number Check | DG invalid serial number | If the DG serial number is empty and does not have any characters. | "}, { 300 , ALARM_ID_HD_INVALID_SERIAL_NUMBER , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. \n\nHD POST: Serial Number Check | HD invalid serial number | If the HD serial number is empty and does not have any characters. | "}, - { 301 , ALARM_ID_DG_RO_PERMEATE_SAMPLE_REMOVE_DIA_CAP , " | | \n\n | | | "}, + { 301 , ALARM_ID_DG_RO_PERMEATE_SAMPLE_REMOVE_DIA_CAP , "DG | Remove Dialysate Cap | Remove the dialysate cap\n - Place the concentrate connection cap on the device.\n- Keep dialysate cap on the device.\n\nCleaning: RO Permeate Sample Connect Cap | Open dialysate cap to be able to collecte the dispensing fluid | RO permeate sample collection state | "}, { 302 , ALARM_ID_DG_CPU_RAM_ERROR , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG Fault: Processor RAM error | DG_CPU RAM error | If the RAM status indicates any of the following: Double bit error, Address decode failure, Address decode logic element failure, Read address parity failure, Write address parity failure. | "}, { 303 , ALARM_ID_DG_DRAIN_PUMP_DIRECTION_FPGA_FAULT , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG Fault: DRP Hall Commutation | DG drain pump direction FPGA fault | If the FPGA indicates an error for a certain period of time. | "}, { 304 , ALARM_ID_HD_INVALID_USAGE_RECORD_CRC , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nHD POST: Usage Rec NVRAM CRC | HD invalid usage reocord | If the CRC of the HD usage record does not match the calculated CRC. | "},