Index: AlarmDefs.h =================================================================== diff -u -re0e4feaeddebc8f7f142502319c01efe2592a5b1 -r633f859a6f9e296b3a15e2446376b4c0a96ae960 --- AlarmDefs.h (.../AlarmDefs.h) (revision e0e4feaeddebc8f7f142502319c01efe2592a5b1) +++ AlarmDefs.h (.../AlarmDefs.h) (revision 633f859a6f9e296b3a15e2446376b4c0a96ae960) @@ -611,7 +611,7 @@ { ALARM_PRIORITY_HIGH, 799, FALSE, TRUE , TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, FALSE, FALSE, ALARM_ID_DG_LOAD_CELLS_INVALID_CALIBRATION, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_INVALID_LOAD_CELL_VALUE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_UV_REACTOR_NOT_HEALTHY, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_DG_FAN_RPM_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, FALSE, TRUE , FALSE, TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, ALARM_ID_DG_FAN_RPM_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_CONCENTRATE_PUMP_FAULT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_CP1_SPEED_CONTROL_ERROR, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_CP2_SPEED_CONTROL_ERROR, 0, ALARM_ID_NO_ALARM },