Index: AlarmDefs.h =================================================================== diff -u -r9d81f449cd0c74b328f16dd249537476ae845991 -rb8f3ed7d03f41829457db6da6db6a912e7b65750 --- AlarmDefs.h (.../AlarmDefs.h) (revision 9d81f449cd0c74b328f16dd249537476ae845991) +++ AlarmDefs.h (.../AlarmDefs.h) (revision b8f3ed7d03f41829457db6da6db6a912e7b65750) @@ -1301,8 +1301,8 @@ { 134 , ALARM_ID_HD_FLUID_LEAK_DETECTED , "HD | Fluid Leak | Fluid leak detected in hemodialysis device.\n- If in treatment, blood pump is off.\n- Identify and correct the cause of the fluid leak.\n- Clean up fluid and dry leak sensor.\n- Once fluid has been cleaned up, Resume/OK treatment. | Blood Stop: Fluid Leak in HD | HD fluid leak detected alarm | If the fluid leak is detected as reported by the fluid leak detector. | "}, { 135 , ALARM_ID_HD_SHOCK , "HD | Blood Pump Stopped: Excessive Movement | Excessive movement of the hemodialysis device has been\ndetected. \n- Treatment is currently paused. \n- When system movement has stablized, \ntreatment will need to be resumed.\n- To resume treatment, press Resume/OK. | Blood Stop: HD Shock Detected | HD experienced large acceleration | If the system experienced excessive shock > 2.5 g. | "}, { 136 , ALARM_ID_DG_SHOCK , "DG | Dialysate Device: Excessive Movement | Excessive movement of the dialysate device has been\ndetected. \n- Treatment is currently paused. \n- When system movement has stablized, \ntreatment will need to be resumed.\n- To resume treatment, press Resume/OK. | Dial. Gen: Shock- Vibration | DG experienced large acceleration | If the system experienced excessive shock > 2.5 g. | "}, - { 137 , ALARM_ID_HD_EXCESSIVE_TILT , "HD | Blood Pump Stopped: Excessive Tilt | Excessive tilt of the hemodialysis device has been detected. \n- Treatment is currently paused. \n- When the device is stationary and level, \ntreatment will need to be resumed.\n- To resume treatment, press Resume/OK. | Blood Stop: HD Tilted | HD tilt exceeds maximum | If the system experienced excessive tilt x and y axis angle > 7 deg for more than 1 sec. | "}, - { 138 , ALARM_ID_DG_EXCESSIVE_TILT , "DG | Dialysate Device: Excessive Tilt | Excessive tilt of the dialysate device has been detected.\n- Treatment is paused.\n- Confirm the device is stationary and level before continuing. | Dial. Gen: Tilt | DG tilt exceeds maximum | If the system experienced excessive tilt x and y axis angle > 7 deg for more than 1 sec. | "}, + { 137 , ALARM_ID_HD_EXCESSIVE_TILT , "HD | Blood Pump Stopped: Excessive Tilt | Excessive tilt of the hemodialysis device has been detected. \n- Treatment is currently paused. \n- When the device is stationary and level, \ntreatment will need to be resumed.\n- To resume treatment, press Resume/OK. | Blood Stop: HD Tilted | HD tilt exceeds maximum | If the system experienced excessive tilt x and y axis angle > 3 deg for more than 1 sec. | "}, + { 138 , ALARM_ID_DG_EXCESSIVE_TILT , "DG | Dialysate Device: Excessive Tilt | Excessive tilt of the dialysate device has been detected.\n- Treatment is paused.\n- Confirm the device is stationary and level before continuing. | Dial. Gen: Tilt | DG tilt exceeds maximum | If the system experienced excessive tilt x and y axis angle > 3 deg for more than 1 sec. | "}, { 139 , ALARM_ID_HD_AC_POWER_LOST , "HD | Power Loss | AC Power has been lost.\n- Check all electrical connections.\n- Press Resume/OK to return to welcome screen. | Loss of AC Power | HD AC power lost alarm | The HD Software shall trigger an alarm if an AC power loss is detected while the HD system is in standby, service, or treatment parameters mode. | "}, { 140 , ALARM_ID_HD_DG_COMM_TIMEOUT , "HD | Communication Error | A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | Loss of DG Comm (Timeout) | Dialysate generator communication timeout | If DG communication timeout is detected for more than 2 seconds. | "}, { 141 , ALARM_ID_HD_AIR_TRAP_FILL_DURING_TREATMENT , "HD | Blood Pump Stopped: Venous Air Detector | A problem was detected with the venous air detector.\n- Confirm arterial access placement and ensure\nconnections are secure.\n- Treatment must be terminated without rinseback. | Blood Stop: Air Trap Fill | Air trap fill timeout during treatment | If the air trap fill timeout during treatment. | "},