Index: firmware/App/Controllers/ConductivitySensors.h =================================================================== diff -u -r025612ad77fe630889a364586de54bffe5262d56 -ref4bc1eaa0f019b678b4d49011a5d23ebe558ad8 --- firmware/App/Controllers/ConductivitySensors.h (.../ConductivitySensors.h) (revision 025612ad77fe630889a364586de54bffe5262d56) +++ firmware/App/Controllers/ConductivitySensors.h (.../ConductivitySensors.h) (revision ef4bc1eaa0f019b678b4d49011a5d23ebe558ad8) @@ -32,17 +32,12 @@ // ********** public definitions ********** -#define MIN_ACID_CONCENTRATE_CONDUCTIVITY 11000 ///< Minimum conductivity for acid concentrate. -#define MAX_ACID_CONCENTRATE_CONDUCTIVITY 11850 ///< Maximum conductivity for acid concentrate. +#define MIN_ACID_CONCENTRATE_CONDUCTIVITY 11062.80 ///< Minimum conductivity for acid concentrate 11645.05 - 5%. +#define MAX_ACID_CONCENTRATE_CONDUCTIVITY 12227.30 ///< Maximum conductivity for acid concentrate 11645.05 + 5%. -#define MIN_BICARB_CONCENTRATE_CONDUCTIVITY 2000 ///< Minimum conductivity for bicarbonate concentrate. -#define MAX_BICARB_CONCENTRATE_CONDUCTIVITY 3000 ///< Maximum conductivity for bicarbonate concentrate. +#define MIN_BICARB_CONCENTRATE_CONDUCTIVITY 13048.14 ///< Minimum conductivity for bicarbonate concentrate 13734.88 - 5%. +#define MAX_BICARB_CONCENTRATE_CONDUCTIVITY 14421.62 ///< Maximum conductivity for bicarbonate concentrate 13734.88 + 5%. -///< Minimum conductivity value for dialysate solution. -#define MIN_DIALYSATE_CONDUCTIVITY ( MIN_ACID_CONCENTRATE_CONDUCTIVITY + MIN_BICARB_CONCENTRATE_CONDUCTIVITY ) -///< Maximum conductivity value for dialysate solution. -#define MAX_DIALYSATE_CONDUCTIVITY ( MAX_ACID_CONCENTRATE_CONDUCTIVITY + MAX_BICARB_CONCENTRATE_CONDUCTIVITY ) - /// Enumeration of conductivity sensors. typedef enum ConductivitySensors {