Index: firmware/App/Controllers/ConductivitySensors.h =================================================================== diff -u -r63215f2052a66ecb3e0355bad0ec99b4db335fff -r9438614daab0edede947e39a9d563fce8f66225e --- firmware/App/Controllers/ConductivitySensors.h (.../ConductivitySensors.h) (revision 63215f2052a66ecb3e0355bad0ec99b4db335fff) +++ firmware/App/Controllers/ConductivitySensors.h (.../ConductivitySensors.h) (revision 9438614daab0edede947e39a9d563fce8f66225e) @@ -45,9 +45,18 @@ */ // ********** public definitions ********** -#define ACID_NORMAL_CONDUCTIVITY 11645.05 ///< Acid normal operation theoretical conductivity. -#define BICARB_NORMAL_CONDUCTIVITY 13734.88 ///< Bicarb normal operation theoretical conductivity. +#define MIN_ACID_CONCENTRATE_CONDUCTIVITY 11000 ///< Minimum conductivity for acid concentrate. +#define MAX_ACID_CONCENTRATE_CONDUCTIVITY 11850 ///< Maximum conductivity for acid concentrate. + +#define MIN_BICARB_CONCENTRATE_CONDUCTIVITY 2000 ///< Minimum conductivity for bicarbonate concentrate. +#define MAX_BICARB_CONCENTRATE_CONDUCTIVITY 3000 ///< Maximum conductivity for bicarbonate concentrate. + +///< Minimum conductivity value for dialysate solution. +#define MIN_DIALYSATE_CONDUCTIVITY ( MIN_ACID_CONCENTRATE_CONDUCTIVITY + MIN_BICARB_CONCENTRATE_CONDUCTIVITY ) +///< Maximum conductivity value for dialysate solution. +#define MAX_DIALYSATE_CONDUCTIVITY ( MAX_ACID_CONCENTRATE_CONDUCTIVITY + MAX_BICARB_CONCENTRATE_CONDUCTIVITY ) + /// Enumeration of conductivity sensors. typedef enum ConductivitySensors {