Index: firmware/App/Controllers/DrainPump.c =================================================================== diff -u -r3e5c16b70fb34da7d58f5cf6efff896fdb3d00c3 -rdef06013312d270b1704787a4473caf5612dd6ac --- firmware/App/Controllers/DrainPump.c (.../DrainPump.c) (revision 3e5c16b70fb34da7d58f5cf6efff896fdb3d00c3) +++ firmware/App/Controllers/DrainPump.c (.../DrainPump.c) (revision def06013312d270b1704787a4473caf5612dd6ac) @@ -367,8 +367,8 @@ isRPMTooHigh |= ( getDrainPumpMeasuredRPM( DRAIN_PUMP_HALL_SNSR_FB ) > MIN_DRAIN_PUMP_RPM ? TRUE : FALSE ); isRPMTooHigh |= ( getDrainPumpMeasuredRPM( DRAIN_PUMP_MAXON_SNSR_FB ) > MIN_DRAIN_PUMP_RPM ? TRUE : FALSE ); - checkPersistentAlarm( ALARM_ID_DG_DRAIN_PUMP_RPM_OUT_OF_RANGE, FALSE, getDrainPumpMeasuredRPM( DRAIN_PUMP_HALL_SNSR_FB ), MAX_ALLOWED_OPEN_LOOP_RPM_OUT_OF_RANGE ); - checkPersistentAlarm( ALARM_ID_DG_DRAIN_PUMP_OFF_FAULT, isRPMTooHigh, getDrainPumpMeasuredRPM( DRAIN_PUMP_HALL_SNSR_FB ), MIN_DRAIN_PUMP_RPM ); + checkPersistentAlarm( ALARM_ID_DG_DRAIN_PUMP_RPM_OUT_OF_RANGE, FALSE, getDrainPumpMeasuredRPM( DRAIN_PUMP_MAXON_SNSR_FB ), MAX_ALLOWED_OPEN_LOOP_RPM_OUT_OF_RANGE ); + checkPersistentAlarm( ALARM_ID_DG_DRAIN_PUMP_OFF_FAULT, isRPMTooHigh, getDrainPumpMeasuredRPM( DRAIN_PUMP_MAXON_SNSR_FB ), MIN_DRAIN_PUMP_RPM ); // If the off fault alarm has become active, trigger the safety shutdown if ( TRUE == isAlarmActive( ALARM_ID_DG_DRAIN_PUMP_OFF_FAULT ) )