Index: firmware/App/Modes/ModeFill.c =================================================================== diff -u -r0cd4bdbd11c1da24df31287cb3f4e3216516ab3f -re4a33b5da88f685fee19f35b9aed2fe1085d56e0 --- firmware/App/Modes/ModeFill.c (.../ModeFill.c) (revision 0cd4bdbd11c1da24df31287cb3f4e3216516ab3f) +++ firmware/App/Modes/ModeFill.c (.../ModeFill.c) (revision e4a33b5da88f685fee19f35b9aed2fe1085d56e0) @@ -409,8 +409,8 @@ F32 const measuredROFlowRate_mL_min = getMeasuredROFlowRate() * ML_PER_LITER; F32 const acidConductivity = getConductivityValue( CONDUCTIVITYSENSORS_CD1_SENSOR ); F32 const dialysateConductivity = getConductivityValue( CONDUCTIVITYSENSORS_CD2_SENSOR ); - BOOL const isAcidConductivityOutOfRange = ( acidConductivity <= MIN_ACID_CONCENTRATE_CONDUCTIVITY ) || ( MAX_ACID_CONCENTRATE_CONDUCTIVITY >= acidConductivity ); - BOOL const isDialysateConductivityOutOfRange = ( dialysateConductivity <= MIN_DIALYSATE_CONDUCTIVITY ) || ( MAX_DIALYSATE_CONDUCTIVITY >= dialysateConductivity ); + BOOL const isAcidConductivityOutOfRange = ( acidConductivity <= MIN_ACID_CONCENTRATE_CONDUCTIVITY ) || ( acidConductivity >= MAX_ACID_CONCENTRATE_CONDUCTIVITY ); + BOOL const isDialysateConductivityOutOfRange = ( dialysateConductivity <= MIN_DIALYSATE_CONDUCTIVITY ) || ( dialysateConductivity >= MAX_DIALYSATE_CONDUCTIVITY ); DG_FILL_MODE_STATE_T result = DG_FILL_MODE_STATE_DELIVER_DIALYSATE; DG_RESERVOIR_ID_T inactiveReservoir = getInactiveReservoir();