Index: firmware/App/Modes/ModeFill.c =================================================================== diff -u -r3325bb3ddb5961167139a9db37b7098083b7b83d -rfb1750ffac99294ef3c797fe30fc16119eb4747d --- firmware/App/Modes/ModeFill.c (.../ModeFill.c) (revision 3325bb3ddb5961167139a9db37b7098083b7b83d) +++ firmware/App/Modes/ModeFill.c (.../ModeFill.c) (revision fb1750ffac99294ef3c797fe30fc16119eb4747d) @@ -614,6 +614,9 @@ setConcentratePumpTargetSpeed( CONCENTRATEPUMPS_CP2_BICARB, BICARB_PUMP_40_ML_PER_MIN ); requestConcentratePumpOn( CONCENTRATEPUMPS_CP2_BICARB ); + // Prior to collecting data for bicarb test, change the CD2 calibration table + setCondcutivitySensorCalTable( CONDUCTIVITYSENSORS_CD2_SENSOR, CAL_DATA_CD2_COND_SENSOR_FILL_BICARB_TEST ); + // State transition concentrateTestStartTime = getMSTimerCount(); result = DG_FILL_MODE_STATE_TEST_BICARB_CONDUCTIVITY; @@ -680,6 +683,10 @@ setROPumpTargetFlowRateLPM( RO_PUMP_800_ML_PER_MIN / MILLILITERS_PER_LITER, TARGET_RO_PRESSURE_PSI ); setConcentratePumpTargetSpeed( CONCENTRATEPUMPS_CP1_ACID, ACID_PUMP_20_ML_PER_MIN ); requestConcentratePumpOn( CONCENTRATEPUMPS_CP1_ACID ); + + // Set back the conductivity of CD2 calibration table to the normal calibration table + setCondcutivitySensorCalTable( CONDUCTIVITYSENSORS_CD2_SENSOR, CAL_DATA_CD2_COND_SENSOR ); + // State transition concentrateTestStartTime = getMSTimerCount(); result = DG_FILL_MODE_STATE_TEST_ACID_CONDUCTIVITY;