Index: firmware/App/Modes/ModeFill.c =================================================================== diff -u -ra6e1a62344275a14e4edde1bdda9974eef840827 -rdbe4474d0bb8524a9a39bac169244d5cf56641bf --- firmware/App/Modes/ModeFill.c (.../ModeFill.c) (revision a6e1a62344275a14e4edde1bdda9974eef840827) +++ firmware/App/Modes/ModeFill.c (.../ModeFill.c) (revision dbe4474d0bb8524a9a39bac169244d5cf56641bf) @@ -322,8 +322,8 @@ { DG_FILL_MODE_STATE_T result = DG_FILL_MODE_STATE_BICARB_PUMP_CHECK; DG_BICARB_CONCENTRATES_RECORD_T bicarb = getBicarbConcentrateCalRecord(); - F32 const measuredROFlowRate_mL_min = getMeasuredROFlowRate() * ML_PER_LITER; - F32 bicarbPumpFlowRate_mL_min = measuredROFlowRate_mL_min * bicarb.bicarbConcentrate[ CAL_DATA_BICARB_CONCENTRATE_1 ].bicarbConcMixRatio + + F32 measuredROFlowRate_mL_min = getMeasuredROFlowRate() * ML_PER_LITER; + F32 bicarbPumpFlowRate_mL_min = measuredROFlowRate_mL_min * bicarb.bicarbConcentrate[ CAL_DATA_BICARB_CONCENTRATE_1 ].bicarbConcMixRatio + CONCENTRATE_PUMP_PRIME_EXTRA_SPEED_ML_MIN; #ifndef DISABLE_DIALYSATE_CHECK F32 const bicarbConductivity = getConductivityValue( CONDUCTIVITYSENSORS_CD2_SENSOR ); @@ -470,11 +470,11 @@ *************************************************************************/ static DG_FILL_MODE_STATE_T handleDeliverDialysateState( void ) { - F32 const measuredROFlowRate_mL_min = getMeasuredROFlowRate() * ML_PER_LITER; - F32 const acidConductivity = getConductivityValue( CONDUCTIVITYSENSORS_CD1_SENSOR ); - F32 const dialysateConductivity = getConductivityValue( CONDUCTIVITYSENSORS_CD2_SENSOR ); - BOOL const isAcidConductivityOutOfRange = ( acidConductivity <= MIN_ACID_CONCENTRATE_CONDUCTIVITY ) || ( acidConductivity >= MAX_ACID_CONCENTRATE_CONDUCTIVITY ); - BOOL const isDialysateConductivityOutOfRange = ( dialysateConductivity <= MIN_DIALYSATE_CONDUCTIVITY ) || ( dialysateConductivity >= MAX_DIALYSATE_CONDUCTIVITY ); + F32 measuredROFlowRate_mL_min = getMeasuredROFlowRate() * ML_PER_LITER; + F32 acidConductivity = getConductivityValue( CONDUCTIVITYSENSORS_CD1_SENSOR ); + F32 dialysateConductivity = getConductivityValue( CONDUCTIVITYSENSORS_CD2_SENSOR ); + BOOL isAcidConductivityOutOfRange = ( acidConductivity <= MIN_ACID_CONCENTRATE_CONDUCTIVITY ) || ( acidConductivity >= MAX_ACID_CONCENTRATE_CONDUCTIVITY ); + BOOL isDialysateConductivityOutOfRange = ( dialysateConductivity <= MIN_DIALYSATE_CONDUCTIVITY ) || ( dialysateConductivity >= MAX_DIALYSATE_CONDUCTIVITY ); DG_FILL_MODE_STATE_T result = DG_FILL_MODE_STATE_DELIVER_DIALYSATE; DG_RESERVOIR_ID_T inactiveReservoir = getInactiveReservoir();