Index: firmware/App/Modes/ModeFill.c =================================================================== diff -u -re12f1b3d41ae133bc0975fcb126e431c2bedda59 -r05120b7716e6c4f8cb06ddd6ec2b29d164148d99 --- firmware/App/Modes/ModeFill.c (.../ModeFill.c) (revision e12f1b3d41ae133bc0975fcb126e431c2bedda59) +++ firmware/App/Modes/ModeFill.c (.../ModeFill.c) (revision 05120b7716e6c4f8cb06ddd6ec2b29d164148d99) @@ -686,7 +686,7 @@ else { // Bicarb test failed, go to fault - SET_ALARM_WITH_1_F32_DATA( ALARM_ID_BICARB_CONDUCTIVITY_OUT_OF_RANGE, averageBicarbConductivity ); + SET_ALARM_WITH_1_F32_DATA( ALARM_ID_DG_BICARB_CONDUCTIVITY_OUT_OF_RANGE, averageBicarbConductivity ); } } @@ -740,7 +740,7 @@ } else { - SET_ALARM_WITH_1_F32_DATA( ALARM_ID_ACID_CONDUCTIVITY_OUT_OF_RANGE, averageAcidConductivity ) + SET_ALARM_WITH_1_F32_DATA( ALARM_ID_DG_ACID_CONDUCTIVITY_OUT_OF_RANGE, averageAcidConductivity ) } if ( pctDiffInConductivity < FIVE_PERCENT_FACTOR ) @@ -937,7 +937,7 @@ if ( ( avgRR > MAX_RO_REJECTION_RATIO_ALLOW ) && ( isROOnlyModeEnabled() != TRUE ) ) { // Fault alarm per PRS 483 - SET_ALARM_WITH_2_F32_DATA( ALARM_ID_RO_REJECTION_RATIO_OUT_OF_RANGE, avgRR, MAX_RO_REJECTION_RATIO_ALLOW ); + SET_ALARM_WITH_2_F32_DATA( ALARM_ID_DG_RO_REJECTION_RATIO_OUT_OF_RANGE, avgRR, MAX_RO_REJECTION_RATIO_ALLOW ); } } } @@ -963,18 +963,18 @@ { setBadAvgConductivityDetectedFlag( TRUE ); // signal idle bad avg conductivity detected setThisFisrtFillFlag( TRUE ); - SET_ALARM_WITH_2_F32_DATA( ALARM_ID_FILL_CONDUCTIVITY_OUT_OF_RANGE, avgBicarbConductivity, bicarbNormalConductivity ); // trigger replace bottles alarm #1 - clearAlarm( ALARM_ID_CREATING_DIALYSATE_PLEASE_WAIT ); // clear this alarm before triggering in case previous fill was bad and still active from before - activateAlarmNoData ( ALARM_ID_CREATING_DIALYSATE_PLEASE_WAIT ); + SET_ALARM_WITH_2_F32_DATA( ALARM_ID_DG_FILL_CONDUCTIVITY_OUT_OF_RANGE, avgBicarbConductivity, bicarbNormalConductivity ); // trigger replace bottles alarm #1 + clearAlarm( ALARM_ID_DG_CREATING_DIALYSATE_PLEASE_WAIT ); // clear this alarm before triggering in case previous fill was bad and still active from before + activateAlarmNoData ( ALARM_ID_DG_CREATING_DIALYSATE_PLEASE_WAIT ); } if ( FALSE == isValueWithinPercentRange( avgAcidConductivity, acidNormalConductivity, FIVE_PERCENT_FACTOR ) ) { setBadAvgConductivityDetectedFlag( TRUE ); // signal idle bad avg conductivity detected setThisFisrtFillFlag( TRUE ); - SET_ALARM_WITH_2_F32_DATA( ALARM_ID_FILL_CONDUCTIVITY_OUT_OF_RANGE, avgAcidConductivity, acidNormalConductivity ); // trigger replace bottles alarm #1 then - clearAlarm( ALARM_ID_CREATING_DIALYSATE_PLEASE_WAIT ); // clear this alarm before triggering in case previous fill was bad and still active from before - activateAlarmNoData ( ALARM_ID_CREATING_DIALYSATE_PLEASE_WAIT ); + SET_ALARM_WITH_2_F32_DATA( ALARM_ID_DG_FILL_CONDUCTIVITY_OUT_OF_RANGE, avgAcidConductivity, acidNormalConductivity ); // trigger replace bottles alarm #1 then + clearAlarm( ALARM_ID_DG_CREATING_DIALYSATE_PLEASE_WAIT ); // clear this alarm before triggering in case previous fill was bad and still active from before + activateAlarmNoData ( ALARM_ID_DG_CREATING_DIALYSATE_PLEASE_WAIT ); } } @@ -1042,14 +1042,14 @@ checkInletWaterPressure(); // Check if any of the alarms are active - if ( ( TRUE == isAlarmActive( ALARM_ID_INLET_WATER_CONDUCTIVITY_IN_LOW_RANGE ) ) || - ( TRUE == isAlarmActive( ALARM_ID_INLET_WATER_CONDUCTIVITY_IN_HIGH_RANGE ) ) || - ( TRUE == isAlarmActive( ALARM_ID_INLET_WATER_TEMPERATURE_IN_HIGH_RANGE ) ) || - ( TRUE == isAlarmActive( ALARM_ID_INLET_WATER_TEMPERATURE_IN_LOW_RANGE ) ) || - ( TRUE == isAlarmActive( ALARM_ID_DG_OUTLET_PRIMARY_CONDUCTIVITY_OUT_OF_RANGE ) ) || - ( TRUE == isAlarmActive( ALARM_ID_RO_REJECTION_RATIO_OUT_OF_RANGE ) ) || - ( TRUE == isAlarmActive( ALARM_ID_INLET_WATER_PRESSURE_IN_HIGH_RANGE ) ) || - ( TRUE == isAlarmActive( ALARM_ID_INLET_WATER_PRESSURE_IN_LOW_RANGE ) ) ) + if ( ( TRUE == isAlarmActive( ALARM_ID_DG_INLET_WATER_CONDUCTIVITY_IN_LOW_RANGE ) ) || + ( TRUE == isAlarmActive( ALARM_ID_DG_INLET_WATER_CONDUCTIVITY_IN_HIGH_RANGE ) ) || + ( TRUE == isAlarmActive( ALARM_ID_DG_INLET_WATER_TEMPERATURE_IN_HIGH_RANGE ) ) || + ( TRUE == isAlarmActive( ALARM_ID_DG_INLET_WATER_TEMPERATURE_IN_LOW_RANGE ) ) || + ( TRUE == isAlarmActive( ALARM_ID_DG_OUTLET_PRIMARY_CONDUCTIVITY_OUT_OF_RANGE ) ) || + ( TRUE == isAlarmActive( ALARM_ID_DG_RO_REJECTION_RATIO_OUT_OF_RANGE ) ) || + ( TRUE == isAlarmActive( ALARM_ID_DG_INLET_WATER_PRESSURE_IN_HIGH_RANGE ) ) || + ( TRUE == isAlarmActive( ALARM_ID_DG_INLET_WATER_PRESSURE_IN_LOW_RANGE ) ) ) { status = TRUE; } @@ -1069,8 +1069,8 @@ { BOOL status = FALSE; - if ( ( TRUE == isAlarmActive( ALARM_ID_BICARB_CONDUCTIVITY_OUT_OF_RANGE ) ) || - ( TRUE == isAlarmActive( ALARM_ID_ACID_CONDUCTIVITY_OUT_OF_RANGE ) ) ) + if ( ( TRUE == isAlarmActive( ALARM_ID_DG_BICARB_CONDUCTIVITY_OUT_OF_RANGE ) ) || + ( TRUE == isAlarmActive( ALARM_ID_DG_ACID_CONDUCTIVITY_OUT_OF_RANGE ) ) ) { status = TRUE; }