Index: firmware/App/Modes/ModeHeatDisinfect.c =================================================================== diff -u -r96b5f969bbba7b44593c85c8ac003be0a2d87151 -r1475b55c0a91327f63a99ee5e40a7b55e0c43a06 --- firmware/App/Modes/ModeHeatDisinfect.c (.../ModeHeatDisinfect.c) (revision 96b5f969bbba7b44593c85c8ac003be0a2d87151) +++ firmware/App/Modes/ModeHeatDisinfect.c (.../ModeHeatDisinfect.c) (revision 1475b55c0a91327f63a99ee5e40a7b55e0c43a06) @@ -67,7 +67,7 @@ #define MAX_FLUSH_CIRC_TEMP_SENSOR_DIFF_C 3.0F ///< Maximum flush circulation temperature difference tolerance in C. #define NUM_OF_TEMP_SENSORS_TO_AVG 3.0F ///< Number of temperature sensors to average to check the difference. #define ACID_PUMP_SPEED_ML_PER_MIN 30.6F ///< Acid concentrate pump speed in mL/min. -// The bicarb pump is 2% faster than the acid pump to create a flow from acid to bicarb line during heat disinfect +// The acid pump is 2% faster than the acid pump to create a flow from acid to bicarb line during heat disinfect #define BICARB_PUMP_SPEED_ML_PER_MIN -30.0F ///< Bicarb concentrate pump speed in mL/min. // Flush and drain R1 and R2